Assembly Replacement Procedures
Disk Drive
4. Remove the disk drive (2) from the carrier assembly.
5. Place the new disk drive onto the carrier assembly and attach with the 4
6. Refer to
7. Locate the Windows Vista License Sticker that came with the replacement
Figure 15-66
Windows License Sticker Rear Panel Location
While the replacement disk drive has the Windows XP operating system loaded onto it, the
license that accompanies it in this kit is for Windows Vista. This is being done within the scope of
our license agreement with Microsoft and is necessary at this time, as the instrument
applications software is not supported or qualified with any operating system other than
Windows XP.
8. Refer to
Keysight N9010A EXA Service Guide
machine screws (1). Torque to 9 in-lbs.
Figure
15-64. Replace the disk drive carrier assembly (2) into the
instrument and tighten the screw to 9 in-lbs.
disk drive and apply it to the rear of the instrument as shown in
15-66, covering the existing license sticker.
Chapter 16, "Post-Repair Procedures."
specified tasks for replacing the A5 Disk Drive assembly.
Figure
and perform all of the
545