RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
If the Reference Assembly calibrator is functioning properly the 4.8 GHz
calibrator signal will measure 4.8 GHz at −28 dBm ± 3 dB as shown in
Figure
4-23. If this level is incorrect, the Reference Assembly is most likely defective.
Reconnect W19 at A9 (1).
Figure 4-23
4.8 GHz Calibrator Signal
178
Keysight N9010A EXA Service Guide