RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Input Attenuator A Power Level Verification
Press Mode Preset, Input/Output, RF Calibrator, 50 MHz, AMPTD,
Attenuation, 0 dB on the analyzer. Refer to
from A9 (1) Output. Measure the 50 MHz calibrator signal on the output of the
attenuator using a functioning Spectrum Analyzer.
Figure 4-13
Cable W11
Keysight N9010A EXA Service Guide
Figure
4-13, remove cable W11
167