Rf Section Description - Keysight Technologies X Series Service Manual

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

RF Section Troubleshooting (Millimeter-Wave Analyzers)

RF Section Description

RF Section Description
This section covers only those optional frequency ranges listed below for the
N9010A Signal Analyzer.
— Option 532, 32 GHz Frequency Range
— Option 544, 44 GHz Frequency Range
The RF input signal can be routed through four different front end signal paths.
These various paths are switched in and out based on where the signal
analyzer center frequency is tuned.
1. Low band RF path is used when the input frequency is ≤ 3600 MHz
2. High band RF path #1 is used when the input frequency is > 3.6 GHz
3. High band RF path #2 is used when the input frequency is > 17 GHz to
4. High band RF Path #3 is used when the input frequency is > 34.4 GHz to
The RF section is designed to convert RF input signals to an intermediate
frequency that is present at the RF output of the A13 Front End assembly. This
intermediate frequency (IF) depends on installed options and the IF path
selected. This chapter will assume the default operation in Spectrum Analyzer
mode. Therefore the IF frequency will be 322.5 MHz.
204
to 17.1 GHz
34.5 GHz
44 GHz
IF Path
10 MHz (standard)
25 MHz (Option B25)
40 MHz (Option B40)
IF
Mode
322.5 MHz
Spectrum Analyzer or IQ Analyzer
322.5 MHz
IQ Analyzer
250 MHz
IQ Analyzer
Keysight N9010A EXA Service Guide

Advertisement

Table of Contents
loading

Table of Contents