Keysight Technologies X Series Service Manual page 98

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Instrument Messages
-221 Settings Conflict Errors
-221
Settings conflict; Gated FFT
is not available while Sweep
Type is set to Swept
-221
Settings conflict; Gated LO
is not available while Sweep
Type is set to FFT
-221
Settings conflict; Gated
Video is not available while
Sweep Type is set to FFT
-221
Settings conflict; Incorrect
RBW for demod. Change RBW
-221
Settings conflict; Ind I/Q
is not available for this
measurement
-221
Settings conflict; Invalid
trace number
-221
Settings conflict; Knob is
not available to modify this
function
-221
Settings conflict; Log Scale
Type is not available when
Detector Mode is Normal
-221
Settings conflict; Log Scale
Type is not available with
Demod View
-221
Settings conflict; Log Scale
Type is only available in
swept measurement
-221
Settings conflict; Log Scale
Type not available when
Segmented Sweep is on
-221
Settings conflict; Log Scale
Type only available when
Frequency Offset = 0 Hz
-221
Settings conflict; Marker 1
Trace Update=off turns off
Signal Track
-221
Settings conflict; Marker
cannot be relative to itself
-221
Settings conflict; Marker
Count is not available when
Gate on
98
The gated FFT function is not available if you have selected the
swept type of sweep. You must be in the FFT sweep type.
The FFT sweep type moves the LO frequency in steps. So the gated
LO function is not available if you have selected FFT sweep.
The gated video function is not available if you have selected the
FFT sweep type.
The Independent I and Q setting is not available for the current
measurement. Only some measurements (initially, only VXA)
support this setting.
The subopcode used to specify the trace number is invalid for this
measurement or query
You should select a specific value for this function. So scrolling
through values with the knob is not allowed.
The logarithmic x-axis scales are not available when you have the
demod view turned on.
Logarithmic scaling can be used when making a swept SA
measurement. It is not available in the SA measurement when you
are using FFT sweeps.
The logarithmic scales are not available while you are using the
segmented sweep capability.
If you have any amount of frequency offset, then you cannot use
logarithmic scaling for the X axis.
Signal Track not available unless the trace containing Marker 1 is
updating
A marker must be set relative to another marker, not to itself.
The marker count function cannot be used while you have gating
turned on.
Keysight N9010A EXA Service Guide