Making Mixer Measurements (Option 089 Only)
High Dynamic Range Swept RF/IF Conversion Loss
Because power meter calibration requires a longer sweep time, you may want
NOTE
to reduce the number of points before pressing
power meter calibration is finished, return the number of points to its original
value and the analyzer will automatically interpolate this calibration.

Perform a Receiver Calibration Over the IF Range

1. Connect the measurement equipment as shown in
Figure 2-17 Connections for Receiver Calibration
2. To calibrate the B-channel over the IF range, press:
INPUT PORTS
Meas
CALIBRATE MENU
Cal
TAKE RCVR CAL SWEEP
Once completed, the analyzer should display −5 dBm (−10 dBm, 8722ES).

Set the Analyzer to the RF Frequency Range

You can find the RF frequency range by using a simple calculation or using the mixer
measurement diagram on the analyzer display.
Using the Calculation
Add the LO frequency to the IF frequency start and stop values. If using an LO frequency
of 1500 MHz, the RF start frequency would be 1.6 GHz (1500 + 100 MHz) and the stop
frequency would be 2.5 GHz (1500 + 1000 MHz).
2-22
B
RECEIVER CAL
TAKE CAL SWEEP
Figure
2-17.
−5
−10
(
x1
x1
. After the
, 8722ES)