Perform A Receiver Calibration Over The If Range; Set The Analyzer To The Rf Frequency Range - Agilent Technologies 8753ET User Manual

Network analyzers
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Because power meter calibration requires a longer sweep time, you may want
NOTE
to reduce the number of points before pressing
the power meter calibration is finished, return the number of points to its
original value and the analyzer will automatically interpolate this
calibration.

Perform a Receiver Calibration Over the IF Range

1. Connect the measurement equipment as shown in
Figure 2-17 Connections for Receiver Calibration
2. To calibrate the B-channel over the IF range, press:
INPUT PORTS
Meas
CALIBRATE MENU
Cal
Once completed, the analyzer should display −5 dBm.

Set the Analyzer to the RF Frequency Range

You can find the RF frequency range by using a simple calculation or using the mixer
measurement diagram on the analyzer display.
Using the Calculation
Add the LO frequency to the IF frequency start and stop values. If using an LO frequency
of 1500 MHz, the RF start frequency would be 1.6 GHz (1500 + 100 MHz) and the stop
frequency would be 2.5 GHz (1500 + 1000 MHz).
High Dynamic Range Swept RF/IF Conversion Loss
B
RECEIVER CAL
Making Mixer Measurements
TAKE CAL SWEEP
Figure
2-17.
−5
TAKE RCVR CAL SWEEP
x1
. After
2-21