Example:
Usage:
:SYSTem:VERSion?
Queries the SCPI version the instrument's command set complies with.
Return values:
<Version>
Example:
Usage:
:SYSTem:COMMunicate:GPIB:LTERminator <LTerminator>
Sets the terminator recognition for remote control via GPIB bus.
Parameters:
<LTerminator>
Example:
:SYSTem:COMMunicate:GPIB:RESource?
Queries the VISA resource string for remote control via the GPIB interface.
To change the GPIB address, use the command
SELF]:ADDRess.
Return values:
<Resource>
User Manual 1176.8674.02 ─ 12
Response: -221, 'Settings conflict', 153,
'Input voltage out of range'
The two returned errors have occurred since the error queue
was last queried.
Query only
string
SYSTem:VERSion
// queries the SCPI version.
Response: "1996"
// the instrument complies with the SCPI version from 1996.
Query only
STANdard | EOI
EOI
The terminator must be sent together with the line message EOI
(End of Line). This setting is recommended for binary block
transmissions where a character could coincidentally have the
value LF (Line Feed) but is not intended as the terminator. This
setting must be selected for block data with undefined length.
STANdard
An LF (Line Feed) is recognized as the terminator regardless of
whether it is sent with or without EOI.
*RST:
n.a. (factory preset: STANdard)
SYSTem:COMMunicate:GPIB:LTERminator EOI
Only a character which is sent simultaneously with the line mes-
sage EOI is accepted as the terminator.
string
Remote control commands
SYSTem subsystem
:SYSTem:COMMunicate:GPIB[:
538