Emerson R&S SGT100A User Manual page 583

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
Error
Code
180
182
183
200
201
202
203
204
241
242
460
461
462
463
464
User Manual 1176.8674.02 ─ 12
Error
Description
Adjustment failed
Adjustment could not be executed
Adjustment data miss-
Adjustment data are missing.
ing
Adjustment data inva-
Adjustment data are invalid and must
lid
be restored.
Cannot access hard-
The data transmission to a module
ware
was unsuccessful.
Hardware revision out
A later version of certain parts of the
of date
instrument is necessary to execute
the function selected.
Cannot access the
An error occurs when writing or read-
EEPROM
ing a EEPROM.
Invalid EEPROM data
Driver initialization
Initialization of a driver fails when
failed
booting the instrument firmware.
No current list
There is no list selected.
Unknown list type
The list type selected is not valid for
specified
the required operation.
For instance, the file extension for
mapping files is *.map. It is not possi-
ble to enter another file extension
when selecting a list.
Cannot open file
The selected file cannot be opened.
Cannot write file
The file cannot be written.
Cannot read file
The file cannot be read.
Filename missing
The required operation cannot be
executed because the file name is not
specified.
Invalid filename exten-
The file extension is not valid for the
sion
required operation.
Error messages and troubleshooting
Device-Specific error messages
Remedy
The adjustment data have to be
generated first by an internal or
external adjustment or to be loa-
ded into the device.
The adjustment data have to be
generated first by an internal or
external adjustment or to be loa-
ded into the instrument.
The adjustment data have to be
generated again by an internal or
external adjustment or to be loa-
ded into the instrument.
The module is not installed, not
properly installed or missing.
The driver does not support the
installed version of a module.
The EEPROM can be defective
and requires replacement.
The driver is not compatible with
the hardware or software config-
uration of the instrument.
To execute the required opera-
tion, a list has to be selected in
the related dialog. If no list is
available, a new list must be cre-
ated.
Check the selected list type.
Check the path and file name.
Check if the file is read-only.
Check if the file contents are
compatible with the file type.
A file name has to be entered
when creating a list.
Check the file extension.
For instance, the file extension
for the mapping files is *.map. It
is not possible to enter another
file extension when storing a list.
583

Advertisement

Table of Contents
loading

Table of Contents