Timeout ← Sensor Settings
For "Filter > Fixed Noise", sets a time limit for the averaging process.
Remote command:
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
Default Aperture Time ← Sensor Settings
The sensor default setting is sufficient. Disable this parameter to specify a user-defined
aperture time per sensor, if, for example, the readings vary.
To obtain stable readings, set the
Remote command:
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
Aperture Time ← Sensor Settings
If "Use Default Aperture Time > Off", defines the acquisition time per sensor.
For example, to obtain a sufficient low average value, set the aperture time exactly to
one modulation period.
Remote command:
:SENSe<ch>[:POWer]:APERture:TIMe
S-Parameter ← Sensor Settings
S-Parameter correction is used to mathematically shift the reference plane to the DUT
by considering the S-parameters for any components connected upstream of the sen-
sor.
The S-Parameter table can be changed with the S-Parameters tool, provided as part of
the free R&S NRP Toolkit software. For more information, refer to the manual of the
connected R&S NRP power sensor.
Remote command:
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect
Enable Logging ← Sensor Settings
Activates recording of R&S NRP power sensor readings in a log file.
There is 1 log file per sensor. The log files are created automatically and filled in con-
tinuously. They are text files with predefined filename SensLog<n>.txt, where <n>
indicates the connected sensor. Log files are stored on the internal memory, in the
directory /var/user/SensorLogging.
Each log file contains the measured value (2 readings when you work with peak sen-
sors), the sensor type, and the measurement time (timestamp). Logged data is not
overwritten. When a new measurement is started, the collected logging data is appen-
ded in the log file.
Check the used disc space regularly and remove log files to maintain storage capacity.
Note: The logging function is intended for measurements with long time intervals. It is
suitable source for data reconstructions if the connection to the sensor was interrupted.
Remote command:
:SENSe<ch>[:POWer]:LOGGing:STATe
User Manual 1176.8674.02 ─ 12
on page 385
Aperture Time
exactly to one modulation period.
on page 382
on page 387
Signal generator settings
RF settings
on page 382
on page 383
on page 383
on page 382