Manual operation:

10.14.13 SOURce:IQ:OUTPut:ENVelope commands

The following remote control commands require software option R&S SGT-K540.
Example: Generating an RF envelope signal and defining the shaping function
// define the RF level and frequency
SOURce1:FREQuency:CW 2143000000
SOURce1:POWer:LEVel:IMMediate:AMPLitude -15
SOURce1:POWer:LEVel:IMMediate:OFFSet 0.5
SOURce1:CORRection:VALue?
// Response: 1
// enable RF envelope generation and define the settings
SOURce1:IQ:OUTPut:ANALog:ENVelope:STATe 1
SOURce1:IQ:OUTPut:ANALog:ENVelope:ADAPtion AUTO
SOURce1:IQ:OUTPut:ANALog:TYPE DIFF
SOURce1:IQ:OUTPut:ANALog:ENVelope:ETRak USER
SOURce1:IQ:OUTPut:ANALog:ENVelope:VREF VCC
SOURce1:IQ:OUTPut:ANALog:ENVelope:POWer:OFFSet?
// Response: 1.5
SOURce1:IQ:OUTPut:ANALog:ENVelope:VPP:MAX 4
SOURce1:IQ:OUTPut:ANALog:ENVelope:GAIN 0
SOURce1:IQ:OUTPut:ANALog:ENVelope:EMF:STATe 1
SOURce1:IQ:OUTPut:ANALog:ENVelope:RIN 50
SOURce1:IQ:OUTPut:ANALog:ENVelope:TERMination GROund
SOURce1:IQ:OUTPut:ANALog:ENVelope:BINPut 1
SOURce1:IQ:OUTPut:ANALog:ENVelope:VCC:OFFSet 2
SOURce1:IQ:OUTPut:ANALog:ENVelope:VCC:MIN 0.5
SOURce1:IQ:OUTPut:ANALog:ENVelope:VCC:MAX 2.5
SOURce1:IQ:OUTPut:ANALog:ENVelope:BIAS 0
SOURce1:IQ:OUTPut:ANALog:ENVelope:OFFSet -2
SOURce1:IQ:OUTPut:ANALog:ENVelope:VOUT:MAX?
// Response: 0.5
SOURce1:IQ:OUTPut:ANALog:ENVelope:VOUT:MIN?
// Response: -1.5
User Manual 1176.8674.02 ─ 12
float
Range:
-0.3V to 0.3V
Increment: 1E-4
*RST:
0
Default unit: V
See
Example"Enabling and configuring the analog output"
on page 481.
See
"Offset (EMF)"
on page 164