:SENSe<ch>[:POWer]:FILTer:TYPE <Type>
Selects the filter mode. The filter length is the multiplier for the time window and thus
directly affects the measurement time.
Parameters:
<Type>
Example:
Manual operation:
:SENSe<ch>[:POWer]:FREQuency <Frequency>
Sets the RF frequency of the signal, if signal source :SENSe<ch>[:POWer]:SOURce
USER is selected.
Parameters:
<Frequency>
Example:
Manual operation:
User Manual 1176.8674.02 ─ 12
SENSe, READ, INITiate and SLISt subsystems
See
"Auto Once"
on page 251
AUTO | USER | NSRatio
AUTO
Automatically selects the filter length, depending on the mea-
sured value. The higher the power, the shorter the filter length,
and vice versa.
USER
Allows you to set the filter length manually. As the filter-length
takes effect as a multiplier of the measurement time, you can
achieve constant measurement times.
NSRatio
Selects the filter length (averaging factor) according to the crite-
rion that the intrinsic noise of the sensor (2 standard deviations)
does not exceed the specified noise content. You can define the
noise content with command
NSRatio.
Note: To avoid long settling times when the power is low, you
can limit the averaging factor limited with the "timeout" parame-
ter (:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe).
*RST:
AUTO
See
Example"Performing a power measurement with a fixed fil-
ter"
on page 377.
See
"Filter"
on page 251
float
*RST:
1 GHz
:SENS1:SOUR USER
Selects user-defined source.
:SENS1:FREQ 2.44GHz
Sets the RF frequency of the source which is 2.44 GHz.
See
"Frequency"
on page 250
Remote control commands
:SENSe<ch>[:POWer]:FILTer:
386