[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP <Step>
The command sets the step width by which the starting power of the carriers in the
defined carrier range will be incremented.
Parameters:
<Step>
Example:
Manual operation:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt] <Start>
The command sets the power for the individual carriers in the defined carrier range. If
the command :BB:ARB:MCAR:EDIT:CARR:POW:STEP is used to define a step width,
the power entered here applies only to the starting carrier. The power of the remaining
carriers is stepped up or down by the power specified in
the :BB:ARB:MCAR:EDIT:CARR:POW:STEP command.
Parameters:
<Start>
Example:
Manual operation:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STARt <Start>
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STOP <Stop>
The command selects the last carrier in the carrier range to which the settings with
the :BB:ARB:MCAR:EDIT:CARR:.. commands shall apply.
Parameters:
<Stop>
User Manual 1176.8674.02 ─ 12
float
Range:
-80 to 80
Increment: 0.01
*RST:
0
Default unit: dB
BB:ARB:MCAR:EDIT:CARR:POW -80dB
sets a power of -80 dB for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:POW:STEP 1 dB
the power is incremented by 1dB for each carrier, i.e. the first
carrier has -80dB, the second -79dB, etc.
See
"Gain Step"
on page 113
See
"Phase Step"
on page 114
float
Range:
-80 to 0
Increment: 0.01
*RST:
0
Default unit: dB
BB:ARB:MCAR:EDIT:CARR:POW -50 dB
sets the power of the carriers in the carrier range to -50 dB.
See
"Gain Start"
on page 113
integer
Range:
0 to 511
*RST:
0
Remote control commands
SOURce subsystem