6.9.1 Important signal parameters

The correct signal processing of input signals in the instrument requires information of
some signal parameters. Among these parameters are sampling rate, crest factor and
signal level, that is expressed as peak or as RMS level value.
Sample rate
The sample rate of the input signal can be defined manually or automatically
retrieved from the input signal.
Crest factor
Test setups can require feeding of external signal to the digital/analog input of the
R&S SGT. If you want to transfer an I/Q signal between two instruments, make
sure that the crest factors match. At the source instrument, transfer an output sig-
nal with a crest factor that matches with the crest factor at the input interfaces of
the receiving instrument.
The crest factor gives the difference in level between the peak level and RMS level
value in dB. The R&S SGT uses the crest factor value for the calculation of power
levels. The general principle is, that the R&S SGT compensates the RMS levels of
the externally supplied signal by the crest factor.
Figure 6-33: Relation between the level settings
In the instrument, level values are represented relative to the full scale.
Fulfilling the prerequisites for a correct signal processing of the externally sup-
plied baseband signal
In the instrument, the crest factor, peak level and RMS level values can be adjusted
in one of the following ways:
Adjusted manually
If you know these values, it is best to enter them directly (see
Estimated automatically by the internal measurement function
Use the function
nal, estimate the peak level and RMS level and calculate the crest factor out of
them.
Tip: For accurate level measurements, set the measurement period to a time value
that is long enough to capture several periods of the input signal.
Determined automatically (digital baseband input only)
The R&S SGT can receive the values from the connected transmitter.
If the external digital signal source is a second Rohde & Schwarz instrument, it sig-
nals the peak level and RMS level of the signal over the digital input interface. To
enable the R&S SGT to receive these values, trigger the
tion.
User Manual 1176.8674.02 ─ 12
External baseband signal - baseband input
Auto Level Set
to trigger the R&S SGT to measure the input sig-
Signal generator settings
Crest
Factor)
DIG IQ Auto Setting
func-
263