Parameters:
<State>
Example:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
Queries the current filter length in filter mode AUTO
FILTer:TYPE)
Return values:
<Auto>
Example:
Usage:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER] <User>
Selects the filter length for
a multiplier for the time window, a constant filter length results in a constant measure-
ment time (see also
achieving stable results"
The R&S NRP power sensors provide different resolutions for setting the filter length,
depending on the used sensor type:
Resolution = 1 for R&S NRPxx power sensors
Resolution = 2
Parameters:
<User>
Example:
User Manual 1176.8674.02 ─ 12
SENSe, READ, INITiate and SLISt subsystems
1 | ON | 0 | OFF
*RST:
0
:SENS1:POW:DISP:PERM:STAT ON
Turns on the permanent view.
See
"Permanent"
on page 250
float
Range:
1 to 65536
:SENS1:FILT:TYPE AUTO
Selects auto filter.
:SENS1:FILT:LENG:AUTO?
Queries the automatically set filter length.
Response: 1024
Query only
See
"Filter Length"
SENS:POW:FILT:TYPE
"About the measuring principle, averaging filter, filter length, and
on page 245).
n
for sensors of the R&S NRP-Zxx family, with n = 1 to 16
float
Range:
1 to 65536
*RST:
1
:SENS1:FILT:TYPE USER
Selects user filter mode.
:SENS1:FILT:LENG 16
Sets a filter length of 16. E.g. using a sensor with 20 ms time
window, the resulting measurement time is 640 ms (2x16x20
ms)
Remote control commands
(:SENSe<ch>[:POWer]:
on page 251
USER. As the filter length works as