:STATus:OPERation:CONDition?
Quieries the content of the CONDition part of the STATus:OPERation register.
This part contains information on the action currently being performed in the instru-
ment. The content is not deleted after being read out because it indicates the current
hardware status.
Return values:
<Condition>
Example:
Usage:
:STATus:OPERation:ENABle <Enable>
Sets the bits of the ENABle part of the STATus:OPERation register. This setting deter-
mines which events of the Status-Event part are forwarded to the sum bit in the status
byte. These events can be used for a service request.
Parameters:
<Enable>
Example:
:STATus:OPERation[:EVENt] <Event>
Queries the content of the EVENt part of the STATus:OPERation register. This part
contains information on the actions performed in the instrument since the last readout.
The content of the EVENt part is deleted after being read out.
Parameters:
<Event>
Example:
:STATus:OPERation:NTRansition <Ntransition>
Sets the bits of the NTRansition part of the STATus:OPERation register. If a bit is set, a
transition from 1 to 0 in the condition part causes an entry to be made in the EVENt
part of the register. The disappearance of an event in the hardware is thus registered,
for example the end of an adjustment.
Parameters:
<Ntransition>
Example:
User Manual 1176.8674.02 ─ 12
string
:STATus:OPERation:CONDition?
Query only
string
:STAT:OPER:ENAB 32767
all events are forwarded to the sum bit of the status byte.
string
:STAT:OPER:EVEN?
queries the STATus:OPERation:EVENt register.
string
:STAT:OPER:NTR 0
a transition from 1 to 0 in the condition part of the Status:Opera-
tion register does not cause an entry to be made in the EVENt
part.