The start line and the number of rows to be filled are defined under "From" and
"Range".
The column to be filled is selected under "Select column to fill". Depending on the
selection here, the default for start, end, and increment value are set. As the settings
are interdependent, a change of one parameter can result in the automatic change of
one or more of the other parameters. The filling of the column with the selected value
settings is started with button "Fill".
The correction list entries are only computed when the "Fill" button is pressed.
From
Sets the start value of the index range.
Remote command:
n.a.
Range
Sets the range for filling the table.
Remote command:
n.a.
Select column to fill
Selects either the frequency or the level column to be filled with the value defined
below.
Remote command:
n.a.
Start value
Sets the start value for the frequency or the level entries.
Remote command:
n.a.
End value
Displays the end value for the frequency or the level entries.
Remote command:
n.a.
Increment value
Sets the increment for the frequency or the level entries.
Remote command:
n.a.
Fill
Fills the selected column in the set range with values, starting with the start value and
using the set increment.
Remote command:
n.a.
User Manual 1176.8674.02 ─ 12
Signal generator settings
RF settings