User Cor. Data - User Correction
Calls the "File Select" menu for selecting and creating a list or the "File Manager".
Remote command:
[:SOURce]:CORRection:CSET:DELete
[:SOURce<hw>]:CORRection:CSET[:SELect]
Edit User Cor. Data
Calls the editor for editing the selected user correction list.
A list consists of any number of frequency/level value pairs. The currently selected list
is displayed.
Each list is saved as a separate file with extension *.uco. The file name and the direc-
tory to which the file is saved are user-selectable. For example,
user correction data for the file ucor.uco.
Note: Save list only after filling both columns (frequency and level), otherwise the
entries are lost.
Figure 6-27: User correction data for file ucor.uco
"Frequency /Hz"
User Manual 1176.8674.02 ─ 12
Enters the frequency to which the level correction value applies.
Note: The "Fill..." function allows you to automatically enter any num-
ber of frequencies with freely selectable range and increment.
Signal generator settings
on page 472
on page 476
Figure 6-27