Depending on the trigger setting, the currently selected segment is
continuously output either immediately or after a trigger event.
Signal generation takes place differently according to the trigger
selected in the "Trigger In" section:
For "Trigger = Auto", output starts at once and the segment is
generated continuously. Trigger events are ignored.
If the segment is changed in "Next Segment", output of the cur-
rent segment is stopped and the new segment is output immedi-
ately, following a system-imposed signal gap (see above).
For "Trigger = Armed_Auto", output starts after the first trigger
event. The segment is then generated continuously. Further trig-
ger events are ignored.
If the segment is changed in "Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
For "Trigger = Retrigger", output starts at once and the segment
is generated continuously, a trigger event causes a restart.
If the segment is changed in "Next Segment", output of the cur-
rent segment is stopped and the new segment is output immedi-
ately, following a system-imposed signal gap (see above).
For "Trigger = Armed_Retrigger", output starts after the first trig-
ger event. The segment is then generated continuously. Further
trigger events cause a restart.
If the segment is changed in "Next Segment", signal output is
stopped and the new segment is not output until a trigger occurs.
For "Trigger = Single", output starts after the first trigger event.
The segment is then generated once. Further trigger events
cause a restart.
If the segment is changed in "Next Segment", signal output is not
stopped. The new segment is not output until a trigger occurs.