Emerson R&S SGT100A User Manual page 129

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
"Next Segment"
User Manual 1176.8674.02 ─ 12
Depending on the trigger setting, the segment selected under "Next
Segment" is output either immediately or after a trigger event.
For "internal Trigger = Auto", output starts at once and the seg-
ment is generated continuously.
If the segment index is changed in "Next Segment", output of the
current segment is stopped and the new segment is output imme-
diately, following a system-imposed signal gap.
For "internal Trigger = Armed_Auto", output starts after the first
trigger event. The segment is then generated continuously.
If the segment index is changed in "Next Segment", output of the
current segment is stopped and the new segment is output imme-
diately, following a system-imposed signal gap.
For "internal Trigger = Armed_Auto", output starts after the first
trigger event. The segment is then generated continuously.
If the segment index is changed in "Next Segment", signal output
is stopped and the new segment is not output until a trigger
occurs.
For "internal Trigger = Single", output starts after the first trigger
event. The segment is then generated once. Further trigger
events cause a restart.
If the segment index is changed in "Next Segment", the new seg-
ment is not output until a trigger occurs.
For "External Trigger = Auto", output starts at once and the seg-
ment is generated continuously.
If the segment index is changed in "Next Segment", output of the
current segment is stopped and the new segment is output imme-
diately, following a system-imposed signal gap.
The subsequent trigger event after the last segment causes the
first segment to be output again.
For "External Trigger = Armed_Auto", output starts after the first
trigger event. The segment is then generated continuously.
If the segment is changed in next segment, signal output is stop-
ped and the new segment is not output until a trigger occurs.
The subsequent trigger event after the last segment causes the
first segment to be output again.
For "External Trigger = Single", output starts after the first trigger
event. The segment is then generated once.
If the segment is changed in "Next Segment", signal output is not
stopped. The new segment is not output until a trigger occurs.
The subsequent trigger event after the last segment causes the
first segment to be output again.