:CALibration:IQModulator:FULL?
Starts the adjustment of the I/Q modulator for the entire frequency range. The I/Q mod-
ulator is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Return values:
<Modulator>
Usage:
Manual operation:
:CALibration:IQModulator:LOCal?
Starts the adjustment of the I/Q modulator for the current frequency. The I/Q modulator
is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
This adjustment is only possible when :OUTPut[:STATe] ON and
[:SOURce]:IQ:STATe ON.
Usage:
Manual operation:
:CALibration:IQModulator:IQModulator[:STATe] <Modulator>
Acitvates/deactivates a separat internal adjustment procedure for the I/Q modulator.
Parameters:
<Modulator>
Manual operation:
:CALibration:LEVel[:MEASure]?
Starts all adjustments which affect the level.
Return values:
<Level>
Usage:
Manual operation:
:CALibration:FREQuency:TEMPerature?
:CALibration:LEVel:TEMPerature?
:CALibration:IQModulator:TEMPerature?
Queries the delta temperature since the last performed adjustment.
Return values:
<Temperature>
Usage:
User Manual 1176.8674.02 ─ 12
1 | ON | 0 | OFF
Query only
See
"I/Q Modulator"
on page 273
Query only
See
"Adjust I/Q Modulator at Current Frequency"
OFF | ON | 1 | 0
See
"I/Q Modulator"
on page 273
1 | ON | 0 | OFF
Query only
See
"Level"
on page 272
string
Query only
Remote control commands
CALibration subsystem
on page 272