Emerson Rosemount CT5100 Reference Manual

Emerson Rosemount CT5100 Reference Manual

Continuous gas analyzer
Hide thumbs Also See for Rosemount CT5100:
Table of Contents

Advertisement

Quick Links

Reference Manual
00809-0100-4512, Rev AA
October 2019
Rosemount
CT5100(Ex)
Continuous Gas Analyzer

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Rosemount CT5100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Emerson Rosemount CT5100

  • Page 1 Reference Manual 00809-0100-4512, Rev AA October 2019 ™ Rosemount CT5100(Ex) Continuous Gas Analyzer...
  • Page 2 Published by Emerson. All possible care has been taken in the preparation of this publication, but Emerson and its agents and distributors accept no liability for any inaccuracies that may be found. This manual reflects the state of the product at the issue date below, but further enhancements while in service may mean that the manual does not reflect your particular system.
  • Page 3 Emerson (Rosemount) designs, manufactures and tests its products to meet many national and international standards. The Rosemount CT5100(Ex) is a sophisticated technical product, and to ensure it continues to operate as designed and within normal specifications it MUST be installed, used, and maintained correctly. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Emerson (Rosemount) products.
  • Page 4 Regulations and standards This article is in accordance with IEC 60079-0: 2011 Clause 30. This article must not be changed amended or removed. Regulations / Standards Description 2014/35/EU The Low Voltage Directive 94/9/EC (until 19 April 2016) The ATEX Directive 2014/34/EU (from 20 April 2016) 2014/30/EU The Electromagnetic Compatibility Directive...
  • Page 5 This equipment meets all requirements of the EMC and Low Voltage directives. Explosive / hazardous area protection Emerson has satisfied the requirements of and complies with IEC, ATEX, and North American regulators for operation of electrical/electronic equipment in hazardous locations .
  • Page 6 Safety and information notices This article is in accordance with IEC 60079-0: 2011 Clause 30. This article must not be changed amended or removed. DANGER WILL CAUSE DEATH Failure to follow this warning will result in death or serious injury to personnel. WARNING DANGER TO PERSONNEL Failure to follow this warning may result in serious injury to personnel.
  • Page 7 Safety Precautions Operators, maintenance personnel, and authorized users must observe the following safety precautions and warnings. Safety precautions are in accordance with IEC 60079-0: 2011 Clause 30. The precautions in this manual MUST NOT be changed amended or removed. All authorized users, installation, operation and maintenance personnel, must observe the following safety precautions and warnings.
  • Page 8 WARNING FIRE AND EXPLOSION In accordance with IEC 60079-0: 2011 Clause 30. Do not open the analyzer's electrical compartment unless the atmosphere in the area is known to be below the ignitable concentration of combustible gases or materials, or unless all equipment within the protected enclosure is de-energized in accordance with NFPA 496 and ISO60079-1/2 and 28.
  • Page 9 WARNING LASER The analyzer contains lasers. Opening the analyzer and attempting to perform adjustments or procedures other than those specified in this manual may result in hazardous optical radiation exposure. All lasers used within the analyzer are Class 1. The emitted laser light is invisible (mid-infrared), and the combined laser powers are sufficiently low at the first accessible aperture that the unprotected eye will not be damaged.
  • Page 10 In accordance with IEC 60079-0: 2011 Clause 30. There are three types of laser that may be included in the Rosemount CT5100(Ex) : Quantum Cascade Lasers (QCLs), Interband Cascade Lasers (ICLs), and diode lasers. The lasers within the analyzer are Class 1. The characteristics of the lasers contained within the analyzer are given in the table below.
  • Page 11 Handle the analyzer with caution during unpacking, installing, maintaining, and transporting to prevent crushing of hands, feet, or other body parts. The analyzer weighs 121 lb. (55 kg). Emerson recommends that a minimum of two people move and lift the analyzer. Wear suitable protective gloves and protective footwear. WARNING EXPLOSION HAZARD / ELECTRIC SHOCK HAZARD In accordance with IEC 60079-0: 2011 Clause 30.
  • Page 12 WARNING EXPLOSION HAZARD In accordance with IEC 60079-0: 2011 Clause 30. After opening the enclosure, do not restore power until the enclosure has been purged for a minimum of 2 minutes 30 seconds at a minimum flow rate of 280 L/m at 1.5 ± 0.5 bar. Failure to observe this warning could cause an explosion or potentially hazardous situation, which if not avoided, may cause death, personal injury, and/or damage to persons and/or property.
  • Page 13 EXPLOSION HAZARD In accordance with IEC 60079-0: 2011 Clause 30. Use only replacement parts and components authorized by Emerson. All replacement parts and components must be certified and approved for use in hazardous areas. Failure to comply will void certification and may cause an explosion or potentially hazardous situation, which if not avoided, may cause death, personal injury, and/or damage to persons and/or property.
  • Page 14 WARNING LOOSE ITEMS Do not place any loose items on top of the system or inside the compartments when doors / covers are open. Confirm that all loose items, tools, and equipment are removed from compartments before closing doors and covers. Failure to observe this warning could cause a potentially hazardous situation, which if not avoided, could result in death or serious injury.
  • Page 15 Some faults can only be repaired by Emerson. Where an item is unserviceable, and no replacement procedure is given in this manual, then the fault must be repaired by Emerson.
  • Page 17: Table Of Contents

    Reference Manual Contents 00809-0100-4512 October 2019 Contents Chapter 1 Introduction.........................21 1.1 Description.............................21 1.2 Customer information........................21 1.3 Safety precautions and conditions for safe use................21 1.4 Qualified personnel........................23 1.5 Software version..........................24 1.6 Glossary and abbreviations......................24 Chapter 2 Theory of operation..................... 27 2.1 Overview............................
  • Page 18 12.12 Replacing the Ethernet connector.................... 143 12.13 Replacing the laser module...................... 144 12.14 Replacing the detector module....................146 12.15 Replacing the O detector......................147 12.16 Replacing the main terminal board...................149 12.17 Replacing the mains power filter....................151 xviii Rosemount CT5100(Ex)
  • Page 19 Reference Manual Contents 00809-0100-4512 October 2019 12.18 Removing the top cover......................151 12.19 Replacing the PT100 gas temperature sensor................152 12.20 Replacing the pressure sensor....................155 12.21 Cleaning the cell mirrors......................156 12.22 Spare parts list..........................162 Chapter 13 Preventative maintenance..................163 13.1 Maintaining the analyzer......................163 13.2 Scheduled maintenance......................
  • Page 20 Contents Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 21: Chapter 1 Introduction

    Cascade Technical Support (cascade.support@emerson.com) or Emerson distribution partners. Further contact details for Emerson can be found on the back page of this manual. Safety precautions and conditions for safe use...
  • Page 22 • The cable glands used in the equipment are only suitable for use in areas with a low risk of mechanical damage and must be suitably protected. Rosemount CT5100(Ex)
  • Page 23: Qualified Personnel

    Reference Manual Introduction 00809-0100-4512 October 2019 • The purge controller provided with this equipment provides alarm signals at various contacts as described in the equipment instructions. The alarms relate to low-flow and loss of purged air supply, and must be connected to the end-user's remote, monitored alarm system.
  • Page 24: Software Version

    Abbreviation Description © Copyright Percent < Less than ° Degree Alternating current ATEX Explosive atmospheres Barg Pressure, in units of bars, above or below atmospheric pressure British Standard Celsius Compressed dry air European Conformity Code of Federal Regulations Rosemount CT5100(Ex)
  • Page 25 Reference Manual Introduction 00809-0100-4512 October 2019 Table 1-1: Abbreviations (continued) Abbreviation Description Continuous Gas Analyzer Methane Carbon dioxide Direct current Degree (temperature) e.g. For example European Community Electromagnetic compatibility European Union Hours Hertz Water Interband Cascade Laser International Electro-technical Commission Inches Ingress protection IPxx...
  • Page 26 Unit of pressure defined as exactly 1/760 of a standard atmosphere UKAS United Kingdom Accreditation Service United States of America Universal serial bus Volt Volt-ampere Volt alternating current Volt direct current Watt WEEE Waste electrical and electronic equipment µm Micro-meter Rosemount CT5100(Ex)
  • Page 27: Theory Of Operation

    Detailed system specifications Laser measurement principle The Rosemount CT5100(Ex) can use up to six lasers to detect and measure gases. Each laser can measure between one and three gases. Inside the laser, which is about the size of a pin head, electrons cascade down a series of quantum wells, producing a photon at each step.
  • Page 28 Theory of operation Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 29: Chapter 3 Description

    The analyzer can be configured to detect and measure up to 10 different gases, depending on the combination of laser modules fitted. System overview A complete Rosemount CT5100(Ex) system consists of a gas handling system, the analyzer, and the associated interconnecting wiring and gas piping. Reference Manual...
  • Page 30 Figure 3-2, the items supplied by Emerson are colored blue; the items supplied by you are colored purple. The green gas handling system may be provided by Emerson or you. Table 3-1 lists the main items of the system.
  • Page 31 Detailed system specifications before entering the analyzer. Detailed characteristics of the analyzer are also given in Table 4-1. Table 3-1: Main Items of the Rosemount CT5100(Ex) Installation Item Name or Supplied by Part number Quantity...
  • Page 32: Gas Flow Through Analyzer

    Description Reference Manual October 2019 00809-0100-4512 Table 3-1: Main Items of the Rosemount CT5100(Ex) Installation (continued) Item Name or Supplied by Part number Quantity Notes description Secondary circuit Customer Customer choice breaker 3.2.1 Explosive safety To prevent explosive vapors from entering the analyzer, the electrical compartment is held at a positive pressure above the surrounding air pressure.
  • Page 33 Reference Manual Description 00809-0100-4512 October 2019 Procedure 1. The gas sample that is to be measured for impurities enters the analyzer through the sample gas input port located on top of the analyzer (see Figure 3-3, Items A and B). 2.
  • Page 34: Connecting The Electrical/Electronic Inputs And Outputs

    Electrical/electronic signal connections to the analyzer are made through three electrical entry points located on the underside of the analyzer, as shown in Figure 3-4. Use the wiring diagram to make the electrical connections as shown in Engineering drawings Figure 3-5. Rosemount CT5100(Ex)
  • Page 35 Reference Manual Description 00809-0100-4512 October 2019 Figure 3-5: Signal Cable Outputs A. To digital modules B. Digital output terminals C. To analog modules D. Analog output terminals E. To status relays F. Status output terminals Note Purge solenoid fitted can be either for 5500-MAN-EX01 for ATEX/IEC applications or 5500- MAN-CD01 for North American applications.
  • Page 36 (www.Pepperl-fuchs.com). It is recommended that these instructions are verified from the Pepperl+Fuchs website to ensure the latest wiring instructions for the purge controller have been followed during installation. The complete wiring diagram used during the manufacture of the Rosemount CT5100(Ex) is included for reference in Engineering drawings.
  • Page 37: Optical Description

    Reference Manual Description 00809-0100-4512 October 2019 • All applicable local and national wiring codes MUST be followed when wiring the system. Also see IEC60079-14. • The power supply to this device shall have a separate customer installed external disconnect. If placed in the hazardous area, it shall be rated for the area it is being installed in.
  • Page 38 To prevent explosive gases from entering the electrical compartment, it is sealed and held at a positive air pressure - above the surrounding air pressure - when the analyzer is switched on. Rosemount CT5100(Ex)
  • Page 39: Chapter 4 Specifications

    Schematic diagrams of the analyzer are shown in Figure 4-1, Figure 4-2, and Figure 4-3. Table 4-1: Physical Characteristics Rosemount CT5100(Ex) Value Comment External dimensions 22.68 x 11.7 x 30.94 in. Length x width x height 575 x 298 x 786 mm Nominal dimensions Weight 121 lb.
  • Page 40 4 to 20 4 or 8 channel outputs, specify on order Communication 10/100 Mbps Ethernet Warm-up time minutes Figure 4-1: Rosemount CT5100(Ex) Dimensions - Front View A. Ventilation B. Lifting eyelet C. User interface D. Purge control panel Rosemount CT5100(Ex)
  • Page 41 Reference Manual Specifications 00809-0100-4512 October 2019 Figure 4-2: Rosemount CT5100(Ex) Dimensions - Top View A. Ventilation B. Lifting eyelet C. Sample return D. Sample inlet Figure 4-3: Rosemount CT5100(Ex) Dimensions - Bottom View A. Atmospheric pressure refer plug B. Earth point C.
  • Page 42: Safety And System Labels And Annotation

    The labels and annotation applied to the analyzer are specified in the table below. Label type Example Location Identification label (including serial Front panel number and model number Fuse identification label 1. Back plate 2. Top right inside of door Rosemount CT5100(Ex)
  • Page 43 Reference Manual Specifications 00809-0100-4512 October 2019 Label type Example Location IECEx and ATEX ratings label Enclosure side panel North America / Canada ratings label Enclosure side panel Reference Manual...
  • Page 44 Reference Manual October 2019 00809-0100-4512 Label type Example Location Universal Ratings Label Enclosure side panel Laser radiation CAUTION label Baseplate Laser module identification label On each laser module housing Intrinsically safe label 1. HMI 2. Intrinsically safe sensor barrier Rosemount CT5100(Ex)
  • Page 45 Reference Manual Specifications 00809-0100-4512 October 2019 Label type Example Location Terminal label (typical layout) Top left inside of door Earth identification label Back plate Manufacturer's label On analysis cell heater block Electrical safety label On inside of electrical compartment door AC Power Supply Danger label 1.
  • Page 46 Specifications Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 47: Chapter 5 Install

    Reference Manual Install 00809-0100-4512 October 2019 Install Site selection ™ The Rosemount CT5100(Ex) has a T3 temperature classification. The user must ensure that no combustible gas concentrations will be present, whether on a continual or occasional basis, which have an ignition temperature below the T3 rating of the analyzer. DANGER FIRE AND EXPLOSION The analyzer's electrical compartment must not be opened unless the atmosphere in the...
  • Page 48: Unpacking

    The analyzer weighs 121 lb. (55 kg) and should always be lifted and moved using suitable lifting/moving equipment. Emerson recommends that a minimum of two people using suitable tools for transportation and lifting are employed.
  • Page 49 The analyzer weighs 121 lb. (55 kg). Emerson recommends that a minimum of two people move and lift the analyzer. Wear suitable protective gloves and protective footwear. Reference Manual...
  • Page 50: Mounting The Analyzer

    7. Ensure that the analyzer is stored in its protective plastic cover until installation. Mounting the analyzer This procedure requires two people to safely move and mount the Rosemount CT5100(Ex). Procedure 1. Ensure that there is free space around the analyzer to allow ventilation of the upper part of the analyzer. Rosemount CT5100(Ex)
  • Page 51 Handle the analyzer with caution during unpacking, installing, maintaining, and transporting to prevent crushing of hands, feet, or other body parts. The analyzer weighs 121 lb. (55 kg). Emerson recommends that a minimum of two people move and lift the analyzer. Wear suitable protective gloves and protective footwear.
  • Page 52 F. Purge compressed air supply port 3. One person should carefully guide the equipment while the other person operates the lifting equipment. 4. Use safety approved and tested lifting equipment to lift the analyzer from the stable platform. Rosemount CT5100(Ex)
  • Page 53 Reference Manual Install 00809-0100-4512 October 2019 5. Mount the analyzer using four M8 (⅜-in.) fasteners to attach the wall mount brackets. The bolts must be positioned in such a way to allow maximum use of all thread length. The installer must ensure that the fasteners used are suitable for the load and surface that the analyzer is mounted on.
  • Page 54 5-2. An Allen key is provided for the grub screws. The protector cap (C) must be retained as it will need to be refitted whenever the analyzer is being shutdown for a prolonged period to seal the purge vent from fluid ingress. Rosemount CT5100(Ex)
  • Page 55 Reference Manual Install 00809-0100-4512 October 2019 Figure 5-4: Mounting Details A. Mounting points Reference Manual...
  • Page 56: Connecting The Electrical/Electronic Inputs And Outputs

    AC power Power is connected to the analyzer instrumentation through the power entry point (B) and the purge controller through the purge power and alarm entry point (A) fitted to the base of the analyzer. Refer to Figure 5-6. Rosemount CT5100(Ex)
  • Page 57 Reference Manual Install 00809-0100-4512 October 2019 Figure 5-6: Power Gland A. Purge power and alarm entry point B. Power entry point Two sources of electrical power are applied to the analyzer: • A purge electrical and alarm supply that is used to power the overpressure facility in the electrical compartment of the analyzer.
  • Page 58 October 2019 00809-0100-4512 Figure 5-7: Power Entry Point Connections A. To mains input fuses B. Mains input filter C. To enclosure earth stud Table 5-1: Mains Input Terminal Terminal Function Sensor system supply (L) Sensor system supply (N) Earth Rosemount CT5100(Ex)
  • Page 59 Reference Manual Install 00809-0100-4512 October 2019 Figure 5-8: Purge Power and Alarm Entry Point Connections A. SV1 B. Power C. Bypass D. To purge solenoid E. To be run separately from other cables looms F. To PT100 sensor Note Purge control system power and alarm contact to be directly wired into purge control unit by end user.
  • Page 60 C. Fuse F1 (mains supply – live) D. Fuse F2 (mains supply – neutral) Table 5-3: Fuse Requirements Fuse Function Rating Schurter part number Live line 110/230 Vac 3.15 A, 250 V, fast 0001.1009 (following mains filter) acting ceramic Rosemount CT5100(Ex)
  • Page 61 Reference Manual Install 00809-0100-4512 October 2019 Table 5-3: Fuse Requirements (continued) Fuse Function Rating Schurter part number Neutral line 110/230 3.15 A, 250 V, fast 0001.1009 Vac (following mains acting ceramic filter) Analyzer 12 Vdc supply 5 A, 250 V, fast acting 0001.1011 rail ceramic...
  • Page 62 Table 5-4: User Connections Terminal Function Digital output 1 Digital output 2 Digital output 3 Digital output 4 Digital output 5 Digital output 6 Digital output 7 Digital output 8 Digital output 9 Digital output 10 Digital output 11 Rosemount CT5100(Ex)
  • Page 63 Reference Manual Install 00809-0100-4512 October 2019 Table 5-4: User Connections (continued) Terminal Function Digital output 12 Analog output 1 Analog output 2 Analog output 3 Analog output 4 Analog output 5 Analog output 6 Analog output 7 Analog output 8 Status output 1 (check function) Status output 2 (maintenance required) Status output 3 (out of specification)
  • Page 64: Pressure Sensor

    NFPA 496. Death, personal injury and/or damage to persons and/or property may result if this is not observed. Rosemount CT5100(Ex)
  • Page 65 5.4.11 Commissioning Once the sensor is fully installed as described above, it should be commissioned in accordance with the commissioning plan agreed between Emerson and the customer. Reference Manual...
  • Page 66 Install Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 67: Controls And Display Controller

    Reference Manual Controls and display controller 00809-0100-4512 October 2019 Controls and display controller Front panel controls and indicators ™ The Rosemount CT5100(Ex) is configured from the two control displays located on the front panel. Refer to Figure 6-1. Figure 6-1: Front Panel A.
  • Page 68 Both external circuit breakers are simple two-pole on/off circuit breakers. Both must be set to On to permit the safe operation of the analyzer. Operation of the analyzer is controlled primarily through the display controller. Refer to Figure 6-3. Rosemount CT5100(Ex)
  • Page 69: Display Controller

    Five LEDs, identified as K1, K2, PSV, Bypass, and PT100, are located on the upper part of the purge control panel. These five LEDs indicate the status of the overpressure air system. Figure 6-4: Purge Control Panel Display controller Operation of the Rosemount CT5100(Ex) is controlled through the six buttons on the display controller: Figure 6-5.
  • Page 70: Gas Sensor Main Screen

    The gas concentrations shown in the following screenshots may be different from those shown in your particular analyzer. The screenshots indicate the functionality of the software, which is the same regardless of the gases or gas concentrations being measured. Rosemount CT5100(Ex)
  • Page 71 Reference Manual Controls and display controller 00809-0100-4512 October 2019 Figure 6-6: Gas Sensor Main Screen The Gas Sensor Main screen displays the gas concentration measurements obtained by the analyzer. In the example shown in Figure 6-6 the gases nitric oxide (NO), nitrogen dioxide ), oxygen (O ), carbon monoxide (CO), and carbon dioxide (CO ) are being measured,...
  • Page 72: Pressure And Temperature Screen

    If, however, the software detects a fault, an error message is displayed. Press STATUS to get further information on the error. Pressure and Temperature screen The Pressure and Temperature screen (Figure 6-8) shows pressure and temperature measurements taken inside the analyzer. Rosemount CT5100(Ex)
  • Page 73: Help System

    Reference Manual Controls and display controller 00809-0100-4512 October 2019 Figure 6-8: Pressure and Temperature Screen The Gas Te reading is the temperature, in °C, of the gas within the analysis cell. The Pres reading is the pressure, in Torr, inside the analysis cell. NOTICE A Torr is a non-SI unit of pressure defined as 1/760 of standard atmospheric pressure and is equal to the fluid pressure of 1 mm of mercury.
  • Page 74: Main Menu

    Figure 6-10: Main Menu BACK button On most of the software screens, the top left-hand button (Figure 6-11) is configured as a BACK button. Press BACK (A) to return to the previous screen. Figure 6-11: BACK Button A. BACK button Rosemount CT5100(Ex)
  • Page 75: Start-Up Procedure

    Reference Manual Start-up procedure 00809-0100-4512 October 2019 Start-up procedure Introduction CAUTION EQUIPMENT DAMAGE Damage to the analyzer may result from a failure to follow this procedure. Always follow the start-up procedure. The analyzer normally operates continuously. It should only be necessary to start up the analyzer under the following circumstances: •...
  • Page 76: Start-Up

    Note The gases shown in the screenshots and the measurements thereof may be different from those shown in your particular Rosemount CT5100(Ex). They indicate the functionality of the software, which is the same regardless of the gases being measured. Note To stop the start-up procedure at any time, set the main circuit breaker to Off.
  • Page 77 Reference Manual Start-up procedure 00809-0100-4512 October 2019 Figure 7-1: Rosemount CT5100(Ex) Gas Inlet and Outlet and Electrical Connections A. Sample gas return port B. Sample gas input port C. Top cover of analyzer (cell compartment) D. Purge air supply E. Laser/electrical compartment F.
  • Page 78 11. Set the main circuit breaker to On. The air pressurization will begin its automatic start-up and begin to purge the air inside the analyzer. Figure 7-2: Purge Control Panel A. K1 LED B. P/SV LED C. PT100 LED 12. Observe the PT100 LED on the purge control panel (Figure 7-2). Rosemount CT5100(Ex)
  • Page 79 Reference Manual Start-up procedure 00809-0100-4512 October 2019 If the PT100 LED (C) illuminates red, the air pressurization system is unserviceable, and the start-up procedure must be halted immediately. To halt the start-up procedure, set the external main circuit breaker to Off. Lock off or tag off the main circuit breaker, turn off the compressed air supply, and report the fault to the maintenance organization.
  • Page 80: Front Panel Controls And Indicators

    After a few seconds, the initial gas concentrations are displayed. The start-up procedure is now complete. Front panel controls and indicators The Rosemount CT5100(Ex) is configured from the control display located on the front panel (Figure 7-4).
  • Page 81 Reference Manual Start-up procedure 00809-0100-4512 October 2019 Figure 7-4: Front Panel A. Display controller B. Purge control panel NOTICE On/Off circuit breakers There are no On/Off switches on the analyzer. The application of electrical power to the analyzer is controlled through a customer provided external circuit breaker. The circuit breaker is a simple two-pole on/off circuit breaker that must be set to On to permit the safe operation of the analyzer.
  • Page 82 7-7) is used to program the operation of the overpressure air system that prevents flammable and explosive gases from entering the electrical compartment of the analyzer. Adjustment of the overpressure air system must only be performed by maintenance personnel. Rosemount CT5100(Ex)
  • Page 83 Reference Manual Start-up procedure 00809-0100-4512 October 2019 Five LEDs, identified as K1, K2, PSV, Bypass, and PT100 are located on the upper part of the purge control panel. These five LEDs indicate the status of the overpressure air system. Figure 7-7: Purge Control Panel (if Applicable) Reference Manual...
  • Page 84 Start-up procedure Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 85: Operating The Analyzer

    Reference Manual Operating the analyzer 00809-0100-4512 October 2019 Operating the analyzer Introduction This section describes the normal operation of the analyzer. NOTICE The gas concentrations shown in the following screenshots may be different from those shown in your particular analyzer. The screenshots indicate the functionality of the software, which is the same regardless of the gases or gas concentrations being measured.
  • Page 86 Operating the analyzer Reference Manual October 2019 00809-0100-4512 Figure 8-1: Gas Sensor Main Screen Figure 8-2: Pressure and Temperature Screen Rosemount CT5100(Ex)
  • Page 87: Chapter 9 Shutdown Procedure

    Reference Manual Shutdown procedure 00809-0100-4512 October 2019 Shutdown procedure Shutdown procedure safety precautions CAUTION EQUIPMENT DAMAGE Damage to the analyzer may result from a failure to follow this procedure. Always follow the shutdown procedure. The analyzer normally operates continuously. It should only be necessary to shut down the analyzer in the following circumstances: •...
  • Page 88: Shutdown Procedure

    (e.g. damage to eyes and skin punctures from debris blown by the high pressure gas or compressed air). 4. Turn off the calibration gas supply to the analyzer. Lock out and tag out the compressed air supply. Rosemount CT5100(Ex)
  • Page 89 Reference Manual Shutdown procedure 00809-0100-4512 October 2019 5. Press MENU on the display controller in either the Gas Sensor Main screen (Figure 9-1) or the Pressure and Temperature screen (Figure 9-2). Figure 9-1: Gas Sensor Main Screen Figure 9-2: Pressure and Temperature Screen The Main menu (Figure 9-3) opens.
  • Page 90 C. Scroll up button D. HELP button E. Scroll down button 8. Use scroll up (C) and scroll down (E) to select SHUTDOWN as shown in Figure 9-4 (B). Then press SEL (select) (B). The Shutdown screen (Figure 9-5) opens. Rosemount CT5100(Ex)
  • Page 91 Reference Manual Shutdown procedure 00809-0100-4512 October 2019 Figure 9-5: Shutdown Screen 9. Press YES. The analyzer shuts down. 10. Set the external secondary circuit breaker to Off. Lock-out and tag-out the secondary circuit breaker. 11. Set the external main circuit breaker to Off. Lock-out and tag-out the main circuit breaker.
  • Page 92 Shutdown procedure Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 93: Chapter 10 Gas Calibration

    Reference Manual Gas calibration 00809-0100-4512 October 2019 Gas calibration 10.1 Required tools The gas concentrations measured by the analyzer can be validated against a known sample gas or calibrated to match it by using the following gas calibration procedure. To calibrate the analyzer, you need the following items: •...
  • Page 94: Main Menu Calibration Routines

    1. To get the Main menu, press the MENU button on the Gas Sensor Main screen (Figure 10-2) or Pressure and Temperature screen Figure 10-3. NOTICE Pressing the BACK button cancels the procedure and returns you to the previous screen. Figure 10-2: Gas Sensor Main Screen Rosemount CT5100(Ex)
  • Page 95 Reference Manual Gas calibration 00809-0100-4512 October 2019 Figure 10-3: Pressure and Temperature Screen 2. The Main menu (Figure 10-4) is used to access the software routines and screens that are used throughout the process. Figure 10-4: Main Menu 3. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. 4.
  • Page 96 7. Use the middle buttons ( and ) to scroll up or down to select the desired gas (e.g., NO). 8. Press the SEL button. 9. To check or change the current offset value, select ZERO OFFSET and press CAL to view. Rosemount CT5100(Ex)
  • Page 97 Reference Manual Gas calibration 00809-0100-4512 October 2019 Figure 10-7: Action Figure 10-8: Calibration 10. Use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to enter the desired offset (in the example above, -0.06ppm). 11. Press OK to confirm. The factory default offset is zero, and under normal circumstances, the offset should never be a positive number.
  • Page 98 1. To get the Main menu, press the MENU button on the Gas Sensor Main screen (Figure 10-2) or Pressure and Temperature screen (Figure 10-3). NOTICE Pressing the BACK button will cancel the procedure and return the user to the previous screen. Rosemount CT5100(Ex)
  • Page 99 Reference Manual Gas calibration 00809-0100-4512 October 2019 2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press the SEL button. Figure 10-11: Main Menu (GAS SERVICE) 3. Use the middle buttons ( and ) to scroll up or down to select PER GAS. Press the SEL button.
  • Page 100 5. Use the middle buttons ( and ) to scroll up or down to select RUN ZERO. Figure 10-14: Action Screen for NO (RUN ZERO) 6. Press the CAL button to calibrate, or the VER button to verify. Figure 10-15 displays as the calibration progresses. Rosemount CT5100(Ex)
  • Page 101 Reference Manual Gas calibration 00809-0100-4512 October 2019 Figure 10-15: Calibration Once complete, the program will automatically open the Result screen (Figure 10-16). Figure 10-16: Result Screen 7. Press OK. The Verify next screen displays. Reference Manual...
  • Page 102 NOTICE Pressing the BACK button will cancel the procedure and return the user to the previous screen. 2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press the SEL button. Rosemount CT5100(Ex)
  • Page 103 Reference Manual Gas calibration 00809-0100-4512 October 2019 Figure 10-18: Main Menu (GAS SERVICE) 3. Use the middle buttons ( and ) to scroll up or down to select PER GAS. Press the SEL button. Figure 10-19: Gas Service (PER GAS) 4.
  • Page 104 00809-0100-4512 Figure 10-20: Select Gas (NO) 5. Use the middle buttons ( and ) to scroll up or down. Select RUN SPAN. Press the CAL button to calibrate, or the VER button to verify. Figure 10-21: Action (SPAN) Rosemount CT5100(Ex)
  • Page 105 Reference Manual Gas calibration 00809-0100-4512 October 2019 6. Using the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons, enter the cylinder concentration (e.g., 20 ppm). Figure 10-22: Calibration Figure 10-22 displays the screen as the calibration progresses. 7. Press OK to confirm. Once complete, the program will automatically open Span Factor For NO (Figure 10-23).
  • Page 106 In order to set the cylinder concentration when a new span gas cylinder is to be used, it is necessary to perform a single gas calibration as described earlier. This concentration will then be stored for future calibrations. Rosemount CT5100(Ex)
  • Page 107 Reference Manual Gas calibration 00809-0100-4512 October 2019 Procedure 1. To get to the Main menu, press the MENU button on the Gas Sensor Main screen (Figure 10-2) or Pressure and Temperature screen (Figure 10-3). NOTICE Pressing the BACK button will cancel the procedure and return the user to the previous screen.
  • Page 108 This completes the process, and the display returns to the Main menu. 10.2.4 Change the automatic calibration settings Procedure 1. To get to the Main menu, press the MENU button on the Gas Sensor Main screen (Figure 10-2) or the Pressure and Temperature screen (Figure 10-3). Rosemount CT5100(Ex)
  • Page 109 Reference Manual Gas calibration 00809-0100-4512 October 2019 NOTICE Pressing the BACK button will cancel the procedure and return the you the previous screen. 2. Use the middle buttons ( and ) to scroll up or down to select GAS SERVICE. Press the SEL button.
  • Page 110 8. To change the time and the units (days, hours, min and sec) between calibrations: Use the LEFT, RIGHT, INCR (increase), and DECR (decrease) buttons to set the number and time interval. Figure 10-33: (Calibration) Period 9. Press the OK button. Rosemount CT5100(Ex)
  • Page 111 Reference Manual Gas calibration 00809-0100-4512 October 2019 10. To adjust the length of time for the span gas to flush the cell: Use the middle buttons ( and ) to scroll up or down select PURGE TIME. Figure 10-34: Calibr param (PURGE TIME) 11.
  • Page 112 Gas calibration Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 113: Troubleshooting And Diagnostics

    Reference Manual Troubleshooting and diagnostics 00809-0100-4512 October 2019 Troubleshooting and diagnostics 11.1 Using the Built-in-Self-Test (BIST) fault diagnostics The analyzer has a Built-In-Self-Test (BIST) function that can be used to perform failure diagnosis of some functions in situations where there is a fault other than a complete failure of the equipment.
  • Page 114 (Figure 11-4) as follows: 11-1, press the MENU button. 5. In the Gas Sensor Main screen, Figure 6. Alternatively, press the MENU button on the Pressure and Temperature screen (Figure 11-2). The Main Menu screen (Figure 11-3) opens. Rosemount CT5100(Ex)
  • Page 115 Reference Manual Troubleshooting and diagnostics 00809-0100-4512 October 2019 Figure 11-3: Main Menu Screen A. Calibration and maintenance options B. SEL text C. SEL button D. Scroll Up button E. Scroll Up arrow F. Scroll Down arrow G. Scroll Down button 7.
  • Page 116: Faults Menu

    The Fault Display screen for the component that is reporting an error will then open. The Fault Display screen will display the parameters of the suspect component. Figure 11-7 shows an example of a Fault Display; the precise data that will be displayed will vary depending upon where the fault has occurred. Rosemount CT5100(Ex)
  • Page 117 Reference Manual Troubleshooting and diagnostics 00809-0100-4512 October 2019 Figure 11-6: Faults Menu A. BACK button B. Fault list C. SEL button D. Scroll Up button E. HELP button F. REFR button G. Scroll Down button Figure 11-7: Fault Display Screen Examine the data shown on the Fault Display to determine if the suspect component has failed or is operating outside of its correct parameters.
  • Page 118: Failure Diagnosis Principles

    In general, if any optical component other than the cell assembly, laser module, and the detector are deemed unserviceable, the analyzer must be repaired by Emerson personnel. The repair, replacement, and alignment of the optical components requires the use of specialized optical test/calibration equipment and procedures to ensure the correct operation of the unit.
  • Page 119: Tools And Test Equipment

    (Figure 11-8). The Diagnostics Menu lists those main components of the Rosemount CT5100(Ex) where problems can be diagnosed using the BIST system. The Diagnostics Menu also enables you to check on the status and, where appropriate, the values of, any of the listed components.
  • Page 120 G. Scroll down button In the example shown in Figure 11-8, LASER 1 has been highlighted. Pressing the SEL button (B) to select the LASER 1 component will open up the Laser 1 diagnostics menu shown in Figure 11-9. Rosemount CT5100(Ex)
  • Page 121 If you do not have the necessary information to determine if the suspect component is faulty, contact Emerson for advice and assistance. If there are more parameters than can be displayed on the available screen space, a PAGE...
  • Page 122: Visual Examination

    00809-0100-4512 11.8 Visual examination If the BIST fails to identify the problem, the next step is a visual examination of the Rosemount CT5100(Ex). Procedure 1. Open the door of the electrical compartment and remove the top cover. Refer to Removing the top cover.
  • Page 123 Reference Manual Troubleshooting and diagnostics 00809-0100-4512 October 2019 3. If PSU is OK, check 12 Vdc fuse (number 3) and replace if necessary. 11.9.4 BEKA and MOXAs do not start up (LEDs not illuminated) Condition All electronics modules are powered up with various on board LEDs. Recommended actions 1.
  • Page 124 Faults page may display MISALIGNMENT. On Gas Sensor 3, weak or no pulses from all lasers with "Misalignment of laser" under Paths in the monitoring tab and possibly "Pulse too noisy" message. Recommended actions 1. Visually check all external optics for dust or damage and clean if necessary. Rosemount CT5100(Ex)
  • Page 125 Reference Manual Troubleshooting and diagnostics 00809-0100-4512 October 2019 2. Ensure cables to detectors are intact. 3. Check all power rails (±12 V and -5 V) are OK and VIGO temp control is working. 4. If not, check power supply ribbon cable from motherboard. 5.
  • Page 126 2. Ensure the 12 V PSU and 12 Vdc fuse are OK; if 12 Vdc fuse is not working, replace it. 3. Ensure connection to BEKA from motherboard is intact; if is not, replace it. 4. Check other outputs (digital and analog) are OK (if available). 5. If all interconnecting cables are OK, replace the motherboard PCB. Rosemount CT5100(Ex)
  • Page 127: Repair And Replacement

    Opening the electrical compartment door To gain access to the electrical compartment and the components mounted on the inside of the door, open the door on the front of the Rosemount CT5100(Ex) as follows: Procedure Using an 8 mm Allen key, loosen the compression screw (B) and cam latch screws (C) to unlock the door (A), then open the door.
  • Page 128: Closing The Electrical Compartment Door

    A. Electrical compartment door B. 1 off compression latch C. 2 off cam latches 12.3 Closing the electrical compartment door The procedure for closing and locking the door to the electrical compartment is the reverse of the opening procedure. Rosemount CT5100(Ex)
  • Page 129: Replacing The Fuses

    CAUTION EQUIPMENT DAMAGE Do not start up or try to operate the Rosemount CT5100(Ex) unless it is physically secure and all electrical and pneumatic connections to the analyzer are in place. Before commencing the start-up process, it is important to ensure that electrical power, sample gas handling facilities, and any calibration gases that are required are available to the analyzer .
  • Page 130 3. No tools are required to remove the fuse. Place a finger on the catch on the upper part of the fuse holder and lift up the upper part of the fuse holder as shown in Figure 12-3. Rosemount CT5100(Ex)
  • Page 131 Reference Manual Repair and replacement 00809-0100-4512 October 2019 4. Fully raise the upper part of the fuse holder (B) as shown in Figure 12-4 and then push out the old fuse (A). Figure 12-3: Replacing the Fuse A. Fuse B. Fuse holder WARNING CORRECT FUSES Fuses must only be replaced with fuses of the same type and rating.
  • Page 132 3.15 A, 250 V, fast 0001.1011 Vac (following mains acting ceramic filter) Analyzer 12 Vdc 5 A, 250 V, fast 0001.1011 supply rail acting ceramic Analyzer 24 Vdc 3.15 A, 250 V, fast 0001.1011 supply rail acting ceramic Rosemount CT5100(Ex)
  • Page 133: Replacing The Tec Board

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 12.5 Replacing the TEC board To replace the TEC board, refer to Figure 12-4 and complete the following steps. Figure 12-4: Replacing the TEC Board A. TEC board B. M3 flat washer C.
  • Page 134 10. Fit the TEC board (A) in position on the backplate (G) in the same orientation. Using the packaging from the replacement TEC board, send the unserviceable TEC board to Emerson for evaluation. 11. Secure the TEC board by fitting the four screws (D) and associated spring washers (C) and flat washers (B) that were retained during the removal procedure.
  • Page 135: Replacing The Peripheral Board

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 12.6 Replacing the peripheral board To replace the peripheral board, refer to Figure 12-5 and complete the following steps. Figure 12-5: Replacing the Peripheral Board A. Peripheral board B. Ribbon cable (to motherboard) C.
  • Page 136 17. Secure the TEC board ribbon cable (D) to the connector of the peripheral board ribbon cable (B) with a cable tie (C). 18. Connect the 12 V power wiring harness (E) to the peripheral board (A). 19. Close the door of the electrical compartment and restart the analyzer. Rosemount CT5100(Ex)
  • Page 137: Replacing The Motherboard

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 12.7 Replacing the motherboard To replace the motherboard, refer to Figure 12-6 and complete the following steps. Figure 12-6: Replacing the Motherboard A. M3 flat washer B. Ribbon cable connector J12 C. M3 spring washer D.
  • Page 138 16. Place the motherboard in position on the backplate and secure by fitting the six screws (D) and the associated spring washers (C) and flat washers (A) retained in Step 13. Torque tighten the screws to 0.6 Nm. Rosemount CT5100(Ex)
  • Page 139: Replacing The Ethernet In/Out Relay Modules

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 17. Connect the HMI display wiring harness (Q) to connector J22 (R). 18. Connect the Ethernet cable (S) to the Ethernet connector (T). 19. Connect the peripheral PCB ribbon cable (K) to connector J7 (J). 20.
  • Page 140: Lower (Main) Din Rail Electronic Components

    DIN rail. Replacing the fuses to replace a fuse. The terminal blocks (I and J) and low-power status relays (A) should not require replacement during the expected life of the equipment. Rosemount CT5100(Ex)
  • Page 141: Replacing A Dc Power Supply

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 Figure 12-8: Lower DIN Rail A. Status relays B. Fuse holder for fuse F4 C. Fuse holder for fuse F3 D. Power supplies E. 10 Amp solid state relay F. Fuse holder for fuse F1 G.
  • Page 142 While continuing to press the lever, remove the DC power supply from the DIN rail. 6. Discard the unserviceable DC power supply. 7. Inspect the replacement DC power supply for damage. 8. Place the replacement DC power supply (A) in its correct location on the DIN rail. Rosemount CT5100(Ex)
  • Page 143: Replace The 10 Amp Solid State Relay

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 9. While pressing the small lever (D) on the underside of the DC power supply to the down position, push the DC power supply onto the DIN rail. 10. Release the small lever and check that the DC power supply is secure. 11.
  • Page 144: Replacing The Laser Module

    Figure 12-10: Replacing the Laser Module A. Base plate B. Pin C. Laser module D. Captive screw E. Ribbon cable F. Captive screw Procedure 1. Shut down the analyzer. Refer to Shutdown procedure. 2. Open the door of the electrical compartment. Rosemount CT5100(Ex)
  • Page 145 Reference Manual Repair and replacement 00809-0100-4512 October 2019 Refer to Figure 12-10. 3. Disconnect the ribbon cable (E). 4. Release the two captive screws (D and F). 5. Remove and discard the unserviceable laser module (C). 6. Inspect the new laser module for any damage. 7.
  • Page 146: Replacing The Detector Module

    F. M3 spring washer G. M3 nut H. Wiring harness Procedure 1. Shut down the analyzer. Refer to Shutdown procedure. 2. Open the door of the electrical compartment. 3. Disconnect the trigger OUT harness (A). Refer to Figure 12-11. Rosemount CT5100(Ex)
  • Page 147: Replacing The O Detector

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 4. Disconnect the wiring harness (H). 5. Remove and retain the four screws (B) nuts (G), plain washers (D) and spring washers (F). 6. Remove and discard the unserviceable detector module (C). 7.
  • Page 148 (type 2) Procedure 1. Shut down the analyzer. Refer to Shutdown procedure. 2. Open the door of the electrical compartment. 3. Disconnect the Bayonet Neill–Concelman (BNC) cable and the Power cable from the detector (D). Refer to Figure 12-13. Rosemount CT5100(Ex)
  • Page 149: Replacing The Main Terminal Board

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 Figure 12-13: Replace O Detector (Type 2) A. M4 socket head cap screw B. M4 spring washer C. M4 plain washer D. O detector sub-assembly E. M4 countersunk screw F. M3 socket head cap screw G.
  • Page 150 9. Secure the main terminal board by fitting the two self-sealing screws (C and D). New self-sealing screws must be used. 10. Connect the wiring harnesses to the main terminal board. 11. Close the door of the electrical compartment and restart the analyzer. Rosemount CT5100(Ex)
  • Page 151: Replacing The Mains Power Filter

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 12.17 Replacing the mains power filter The mains power filter is located on the underside of the main terminal board. Its purpose is to protect the analyzer against spikes and surges in the mains power. Procedure 1.
  • Page 152: Replacing The Pt100 Gas Temperature Sensor

    Replacing the PT100 gas temperature sensor Note Approval for the PT100 temperature sensor as a simple device: To permit connection of a suitable temperature sensor, the following intrinsically safe (IS) parameters have been declared for the connections emerging from the purged enclosure. Rosemount CT5100(Ex)
  • Page 153 Reference Manual Repair and replacement 00809-0100-4512 October 2019 Parameter Voltage U 12.4 V Current I 17.4 mA Power P 54 mW Capacitance C Inductance L The capacitance and either the inductance or L/R of the load connected to hazardous area terminals of the apparatus must not exceed the following values for Group IIC: Capacitance Inductance...
  • Page 154 13. Use one spanner to hold the main body to prevent it turning while using a torque wrench with crow foot attachment to tighten cap (G). 14. Route the gas temperature sensor wiring harness (B), as noted down in Step 3, to the intrinsically safe barrier. Rosemount CT5100(Ex)
  • Page 155: Replacing The Pressure Sensor

    Reference Manual Repair and replacement 00809-0100-4512 October 2019 15. Secure the gas temperature sensor wiring harness with cable ties and cable clips in the locations that were noted down in Step 16. Fit the top cover and restart the analyzer. 12.20 Replacing the pressure sensor Figure 12-17: Replacing the Pressure Sensor...
  • Page 156: Cleaning The Cell Mirrors

    Equipment materials The following equipment and materials are required when cleaning the cell mirrors: • Spectroscopic grade methanol or ethanol • Optical grade lens tissue • Invertible air duster • Hex head keys, sizes 1.5 mm to 4 mm Rosemount CT5100(Ex)
  • Page 157 Reference Manual Repair and replacement 00809-0100-4512 October 2019 12.21.2 Removing the cell lid Figure 12-18: Removing Analysis Cell Lid A. Cell lid B. Cell lid fasteners C. Cell lid insulation D. Tube fitting nut E. Bracket F. Pressure sensor G. Bracket fastener H.
  • Page 158 12.21.3 Cleaning the cell mirrors with air duster Emerson recommends initially cleaning the cell mirrors with an invertible air duster to remove any particulates or dust that may have contaminated the surface. Following cleaning with the air duster, check the pulses to see if any improvement has been made.
  • Page 159 Reference Manual Repair and replacement 00809-0100-4512 October 2019 Figure 12-19: Optical Grade Lens Tissue 2. Apply a small amount (one or two drops) of spectroscopic grade methanol to the front of the tissue loop. 3. To prevent scratching the mirrors do not apply any direct pressure to the mirrors (Figure 12-20).
  • Page 160 7. Allow the cell temperature to stabilize and check the pulses in the gas sensor software. If the pulses have returned to a usable signal size proceed to operate the analyzer. Note If pulses have not returned to a usable signal size, contact Emerson for assistance. Rosemount CT5100(Ex)
  • Page 161 Reference Manual Repair and replacement 00809-0100-4512 October 2019 12.21.5 Assembling analysis cell Figure 12-22: Assembling Analysis Cell A. Cell lid B. Cell lid fasteners C. Cell lid insulation D. Tube fitting nut E. Bracket F. Pressure sensor G. Bracket fastener H.
  • Page 162: Spare Parts List

    Heater 120 V, certified P-6001-00015 SL Blocktherm DKA 80 W heater block Power supply 12 V P-6001-00017 72 W switch mode DIN rail panel mount power Power supply 24 V P-6001-00018 90 W switch mode DIN rail panel mount power Rosemount CT5100(Ex)
  • Page 163: Chapter 13 Preventative Maintenance

    Reference Manual Preventative maintenance 00809-0100-4512 October 2019 Preventative maintenance 13.1 Maintaining the analyzer WARNING MAINTENANCE / MODIFICATIONS On completion of any maintenance and/or modifications, verify: — All tools and equipment are removed. — No contamination (water or dust) is in the compartments. —...
  • Page 164 Preventative maintenance Reference Manual October 2019 00809-0100-4512 WARNING SAFE OPERATING STATE ™ On completion of any maintenance, ensure that the Rosemount CT5100(Ex) is in a safe and correct operating state. Rosemount CT5100(Ex)
  • Page 165: Atex/Iecex Assessment Report Summary

    FK9 4TZ 14.3 Manufacturer's name and address As applicant. 14.4 Trademark 14.5 Product name/model number Rosemount CT5100(Ex) Continuous Gas Analyzer 14.6 Rating 110 V to 120 V, 60 Hz 220 V to 240 V, 50 Hz 1100 W Reference Manual...
  • Page 166: Assessment Standards

    Certification code Class I, Division 2, Groups A, B, C, D; T3. Tamb-20 °C to +55 °C Marking class CLASS– 2258-82 Certification code Class I, Division 2, Groups A, B, C, D; T3. Tamb-20 °C to +55 °C Rosemount CT5100(Ex)
  • Page 167 Reference Manual ATEX/IECEX assessment report summary 00809-0100-4512 October 2019 Table 14-4: Common Markings Details Model number Rosemount CT5100(Ex) Continuous Gas Analyzer Manufacturer's name Cascade Technologies Ltd Manufacturer's address Glendevon House Castle Business Park Stirling FK9 4TZ Ambient range -4 to 131 °F -20 to 55 °C...
  • Page 168 été isolée. Il convient de retirer la batterie s’il faut que l'enveloppe reste sans protection ex pendant une durée prolongée. Type of protective gas Instrument quality air or inert gas Minimum overpressure 0.5 mbar Maximum overpressure 10 mbar Rosemount CT5100(Ex)
  • Page 169: Conditions Of Certification/Special Conditions For Safe Use

    25 L/min for leakage compensation. • When using the Rosemount CT5100(Ex) variant fitted with the Elmess gas cell heater certified as BVS 14ATEXE155U and IECEx BVS 14.0106U, the equipment top compartment should be cooled with a source of compressed air at a minimum flow rate of 280 L/min.
  • Page 170: Conditions Of Certification For The Replacement Of Simple Devices

    NFPA496 Clause 4.12.5. • When installing conduit for power and data connections the end-user must select suitably certified conduit. 14.10 Conditions of certification for the replacement of simple devices Approval for the PT100 temperature sensor as a simple device. Rosemount CT5100(Ex)
  • Page 171 Reference Manual ATEX/IECEX assessment report summary 00809-0100-4512 October 2019 • To permit connection of a suitable temperature sensor, the following intrinsically safe (IS) parameters have been declared. • For the connections emerging from the purged enclosure: Parameter Voltage U 12.4 V Current I 17.4 mA Power P...
  • Page 172 ATEX/IECEX assessment report summary Reference Manual October 2019 00809-0100-4512 Rosemount CT5100(Ex)
  • Page 173: Appendix A Engineering Drawings

    CT5100(Ex) analyzer to assist with troubleshooting faults. These diagrams may be used to locate the position of a wiring connector should it become disconnected. Table A-1: List of Engineering Drawings Drawing number Description W-2000-0034 (Sheets 1-5) Rosemount CT5100(Ex) heated Reference Manual...
  • Page 174 23/MAR/2016 SHEET THIS DRAWING AND DATA CONTAINED WITHIN IT ARE THE SOLE CASCADE TECHNOLOGIES LTD DESCRIPTION: PROPERTY OF CASCADE TECHNOLOGIES LTD, PART OF EMERSON DO NOT SCALE GLENDEVON HOUSE-CASTLE BUSINESS PARK BALLOON I.D. NO. PROCESS MANAGEMENT AND CANNOT BE COPIED,...
  • Page 175 23/MAR/2016 SHEET THIS DRAWING AND DATA CONTAINED WITHIN IT ARE THE SOLE CASCADE TECHNOLOGIES LTD DESCRIPTION: PROPERTY OF CASCADE TECHNOLOGIES LTD, PART OF EMERSON DO NOT SCALE GLENDEVON HOUSE-CASTLE BUSINESS PARK BALLOON I.D. NO. PROCESS MANAGEMENT AND CANNOT BE COPIED,...
  • Page 176 23/MAR/2016 SHEET THIS DRAWING AND DATA CONTAINED WITHIN IT ARE THE SOLE CASCADE TECHNOLOGIES LTD DESCRIPTION: PROPERTY OF CASCADE TECHNOLOGIES LTD, PART OF EMERSON DO NOT SCALE GLENDEVON HOUSE-CASTLE BUSINESS PARK BALLOON I.D. NO. PROCESS MANAGEMENT AND CANNOT BE COPIED,...
  • Page 177 23/MAR/2016 SHEET THIS DRAWING AND DATA CONTAINED WITHIN IT ARE THE SOLE CASCADE TECHNOLOGIES LTD DESCRIPTION: PROPERTY OF CASCADE TECHNOLOGIES LTD, PART OF EMERSON DO NOT SCALE GLENDEVON HOUSE-CASTLE BUSINESS PARK BALLOON I.D. NO. PROCESS MANAGEMENT AND CANNOT BE COPIED,...
  • Page 178 23/MAR/2016 SHEET THIS DRAWING AND DATA CONTAINED WITHIN IT ARE THE SOLE CASCADE TECHNOLOGIES LTD DESCRIPTION: PROPERTY OF CASCADE TECHNOLOGIES LTD, PART OF EMERSON DO NOT SCALE GLENDEVON HOUSE-CASTLE BUSINESS PARK BALLOON I.D. NO. PROCESS MANAGEMENT AND CANNOT BE COPIED,...
  • Page 179 2019 Emerson. All rights reserved. twitter.com/rosemount_news The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a Facebook.com/Rosemount mark of one of the Emerson family of companies. All other marks are the property of their youtube.com/RosemountMeasurement respective owners.

This manual is also suitable for:

Rosemount ct5100 ex

Table of Contents