5. Enter a program number.
6. Press ENTER key.
The SET DISPLAY will then return to the HOME screen, and that program number will appear
in the PROG # field as confirmation:
7. Record saved information on the Program Log Pad (supplied).
Note To prevent accidental erasure of established programs, if the entered program number has
been previously assigned, the following message will appear:
To review stored programs before overwriting, refer to your Program Log Pad or press RECALL. If you
press RECALL, you will have to go back and re-establish the run parameters before you can store them
to memory.

Performing a Programmed Run

Read the Safety Information Page at the front of this manual.
1. Press RECALL .
The SET DISPLAY will change to display a program number flashing in the PROG # field.
2. Input the desired program number, then press ENTER key.
The SET DISPLAY and options indicators change to reflect the run parameters that were saved to
that program number.
3. Prepare the rotor according to the rotor manual.
4. Open the chamber door and install the rotor, making sure that the rotor is fully seated and locked
to the drive spindle. Install the rotor cover, if any, then close the chamber door.
Note Rotor names are saved within the program. If the rotor is changed by selecting a different
program this will be indicated in the SET DISPLAY. Be sure that the installed rotor is correct for
the recalled program. Make sure that the SET rotor identifies the installed rotor.
5. If using logging software, make sure it is running, then press START key.
The rotor will accelerate according to the recalled parameters, and continue until run termination. At
termination, the rotor will decelerate to a stop. After it stops "End" will appear in the SPEED display
and remains there until the door is opened or the START key will be pressed.
Note If you select a program number that has not been used to save a program, a warning message
will be displayed. In this case no program will be recalled.
Thermo Scientific
4
Operation
Programmed Operation
RC 12BP+
4-7