Common Command Reference
The following reference discusses the HP E1313/E1413 IEEE-488.2 Common
Commands.
Calibration Command. The calibration command causes the Channel Calibration
function to be performed for every module channel. The Channel Calibration
function includes calibration of A/D Offset, and Gain and Offset for all 64 channels.
This calibration is accomplished using internal calibration references. The
command causes the module to calibrate A/D offset and gain, and all channel
offsets. This may take many minutes to complete. The actual time it will take your
HP E1313/E1413 to complete
performs literally hundreds of measurements of the internal calibration sources for
each channel and must allow 17 time constants of settling wait each time a filtered
channel's calibrations source value is changed. The
very sophisticated and results in an extremely well calibrated module.
To perform Channel Calibration on multiple HP E1313/E1413s, use the
command (see
CALibration:SETup
•
Returned Value:
Value
0
-1
The C-SCPI type returned is int16.
• CALibration:STORe ADC
into non-volatile memory.
CALibration:SETup
•
Executing this command does not alter the module's programmed state (function,
range, etc.), but it does clear bit 13 in the Questionable Data Group (see Table 5-4
on page 259).
•
This command will cause Error +3000, "illegal while initiated" if trigger is initiated.
•
Related Commands:
CALibration:SETup, CALibration:SETup?
CALibration:STORe ADC
Common Command Reference
depends on the mix of SCPs installed.
*CAL?
*CAL?
CALibration:SETup
Meaning
Cal OK
Cal Error
See Error Messages in Appendix B
stores the calibration constants for
HP E1313/E1413 Command Reference
*CAL?
*CAL?
procedure is internally
on page 179 for details).
Further Action
None
Query the Error Queue
(SYSTem:ERRor?)
and
*CAL?
,
275