Usage
DIAGnostic:FLOor:DUMP
Comments
DIAGnostic:INTerrupt[:LINe]
Parameters
Comments
Usage
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
•
Once a channel's autorange floor is set by
reset by another
DIAGnostic:FLOor
•
A channel with an autorange floor can be manually ranged below the floor
(
[SENSe:]FUNCtion:...
the autorange floor setting is still in effect.
•
Related Commands:
DIAGnostic:FLOor:DUMP
•
*RST Condition:
DIAGnostic:FLOor 0,(@100:163)
DIAG:FLO 1,(@100,124)
DIAGnostic:FLOor:DUMP
the FIFO.
•
The format of the values returned from the FIFO with a
depends on the format chosen by the
•
Related Commands:
DIAGnostic:FLOor[:CONFigure]
commands
[SENSe:]FUNCtion:...
•
*RST Condition:
DIAGnostic:FLOor:DUMP
64 channels.
DIAGnostic:INTerrupt[:LINe] <
module will use.
Parameter
Parameter
Name
Type
intr_line
numeric (int16)
<
>
•
Related Commands:
DIAGnostic:INTerrupt[:LINe]?
•
Power-on and *RST Condition:
DIAG:INT:LIN 5
DIAGnostic:FLOor
command or by a
commands). When the channel is returned to autorange,
,
[SENSe:]FUNCtion:...
Channel 0 and 24 can autorange no lower
than 1 V range.
sends the autorange floor settings for all 64 channels to
FORMat:DATA
places 0 values in the FIFO for all
intr_line
>
sets the VXIbus interrupt line the
0 through 7
DIAGnostic:INTerrupt:LINe 1
Module will interrupt on VXIbus interrupt
line 5.
HP E1313/E1413 Command Reference
DIAGnostic
, it remains until
*RST
command.
commands.
command
SENSe:DATA?
command.
,
,
FORMat[:DATA]
Range of
Default
Values
Units
none