DIAGnostic

DIAGnostic:CUSTom:LINear

Parameters
Comments
Usage
190
HP E1313/E1413 Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
DIAGnostic:CUSTom:LINear <
downloads a custom linear Engineering Unit Conversion table (in
the HP E1313/E1413. Contact your Hewlett-Packard System Engineer for more
information on Custom Engineering Unit Conversion for your application.
Parameter
Parameter
Name
Type
<
table_range
>
numeric (float32)
<
table_block
>
definite length block
data
<
ch_list
>
channel list (string)
• <
table_block
is a block of 8 bytes that define 4, 16-bit values. SCPI requires that
>
table_block
include the definite length block data header. C-SCPI adds the
<
>
header for you.
• <
table_range
>
specifies the range of voltage that the table covers
(from –
<
table_range
>
be within 5% of one of the nominal values from the table above.
• <
ch_list
>
specifies which channels may use this custom EU table.
Related Commands:
[SENSe:]FUNCtion:CUSTom
*RST Condition: All custom EU tables erased.
Program puts table constants into array table_block.
table_block
DIAG:CUST:LIN
SENS:FUNC:CUST:LIN 1,1,(@116:123)
INITiate then TRIGger module.
table_range
table_block
>,<
Range of
.015625 | .03125 | .0625 | .125 | .25 | .5 |
1 | 2 | 4 | 8 | 16 | 32 | 64
See Comments below.
100 - 163
to +
<
table_range
>
). The value you specify must
,(@116:123) Send table to HP E1413 for channels 16 - 23.
Link custom EU with channels 16 - 23.
ch_list
>,(@<
>)
<
table_block
Default
Values
Units
volts
none
none