parameter: The HP E1413 has five ranges: .0625 VDC, .25 VDC,
1 VDC, 4 VDC, and 16 VDC. To select a range, simply specify the range value
(for example, 4 selects the 4 VDC range). If you specify a value larger than one of
the first four ranges, the HP E1413 selects the next higher range (for example, 4.1
selects the 16 VDC range). Specifying a value larger than 16 generates Error -222,
"Data out of range" . Specifying
selects autorange. The default range (no range parameter specified) is
AUTO
autorange. Autorange is not allowed while
Error +3072, "Autorange not allowed with SENSE:FILTER on" will be generated at
INITiate:IMMediate
time if the filter is
<
range
>
may be specified in millivollts (mv).
If you are using amplifier SCPs, you should set them first and keep their settings
in mind when specifying a range setting. For instance, if your expected signal
voltage is to be approximately .1 VDC and the amplifier SCP for that channel has
a gain of 8, you must set
condition will exist.
The
<
type
> EEXTended
The
*CAL?
command calibrates temperature channels based on Sense Amplifier
SCP setup at the time of execution. If SCP settings are changed, those channels
are no longer calibrated.
Related Commands:
DIAGnostic:CUSTom:PIECewise
[SENSe:]REFerence
,
[SENSe:]REFerence:TEMPerature
*RST Condition: All custom EU tables are erased.
Program must put table constants into array table_block.
DIAG:CUST:PIEC 1,table_block,(@100:107)Send characterized thermocouple table for
SENS:FUNC:CUST:TC N,.25,(@100:107) Link custom thermocouple EU with
SENS:REF RTD,92,(@120)
Include these channels in a scan list (REF channel first).
INITiate then TRIGger module.
selects the lowest range (.0625 VDC). Specifying
0
SENSe:FILTer:LPASs:STATe
ON
and any channel specifies autorange.
range
no lower than 1 VDC or an input out-of-range
<
>
applies to E type thermocouples at 800°C and above.
must be executed again.
*CAL?
use by channels 0 - 7.
channels 0 - 7; use reference temperature
compensation for N type wire.
Designate a channel to measure the
reference junction temperature.
HP E1313/E1413 Command Reference