Unshielded signal wiring is very common in Data Acquisition applications.
While this worked well for low speed integrating A/D measurements and/or for
measuring high level signals, it does not work for high speed sampling A/Ds,
particularly when measuring low level signals like thermocouples or strain gage
bridge outputs. Unshielded wiring will pick up environmental noise, causing
measurement errors. Shielded, twisted-pair signal wiring, although it is
expensive, is required for these measurements unless an even more expensive
"amplifier at the signal source" or individual A/D at the source is used.
Generally, the shield should be connected to ground at the DUT and left
open at the HP E1313/E1413. Floating DUTs or transducers are an
exception. Connect the shield to HP E1313/E1413 GND or GRD terminals
for this case, whichever gives the best performance. This will usually be the
GND terminal. A single point shield to ground connection is required to
prevent ground loops. This point should be as near to the noise source as
possible and this is usually at the DUT.
The following lists some recommended wiring techniques.
1. Use individually shielded, twisted-pair wiring for each channel.
2. Connect the shield of each wiring pair to the corresponding Guard (G)
terminal on the terminal module.
3. The terminal module is shipped with the Ground-Guard (GND-GRD)
shorting jumper installed for each channel. These may be left
installed or removed, dependent on the following conditions:
a. Grounded Transducer with shield connected to ground at the
transducer: Low frequency ground loops (DC and/or 50/60 Hz)
can result if the shield is also grounded at the terminal module end.
To prevent this, remove the GND-GRD jumper for that channel.
b. Floating Transducer with shield connected to the transducer
at the source: In this case, the best performance will most likely
be achieved by leaving the GND-GRD jumper in place.
4. In general, the GND-GRD jumper can be left in place unless it is
necessary to break low frequency (below 1 kHz) ground loops.
Wiring and Noise Reduction Methods
Appendix F
401