Internal Timer Based Scans - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

Internal Timer
Based Scans
116
Understanding the HP E1313/E1413
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Card 2
ROUT:SEQ:DEF LIST<
n
>,(@101,103,105,107) <n> can be 1 - 4.
ROUT:SCAN LIST<
n
>
SAMP:TIM LIST<
n>,<t
>
c
TRIG:SOUR IMM
source
ARM:SOUR <
>
INIT:CONT ON
Starting
ARM event, either hardware or software
Acquisition starts
Stopping
INIT:CONT OFF
In this case both cards are set up to be paced by their trigger timers. Card 1 is
also set up to source a trigger on one of the VXIbus TTLTRG lines and card 2
is set up to be triggered by that line. This means that triggering card 1 will start
both cards.
t
p
t
c
Card 1
Master
ARM
Card 2
Slave
t
is the sample time between channels.
c
t
is the pacing time between channel lists.
p
The large arrow is the pacing Trigger Timer. The smaller arrows are sample
triggers generated by the HP E1313/E1413 when stepping through the
channel list. The first large arrow is the arming event.
<t
> can be 10µs to 32.768ms.
c
<source> can be BUS, EXT,
HOLD, IMM, or TTLTrg0 -
TTLTrg7.
To both cards, stops at end of
scan list.
=Trigger Count
Channel Number
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1413c

Table of Contents