HP E1313A Manual page 245

High speed a/d module
Table of Contents

Advertisement

[SENSe:]
Usage
244
HP E1313/E1413 Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
sub_type
parameter: values of 85 and 92 differentiate between 100 Ohm
<
>
(@ 0°C) RTDs with temperature coefficients of 0.00385 and and 0.00392
Ohm/Ohm/°C respectively. The
refer to thermistors that match the Omega 44000 series temperature response
curve. These 44000 series thermistors are selected to match the curve within 0.1
or 0.2 °C. For thermistors
The
<
sub_type
> EEXTended
CUSTom
is pre-defined as Type K, with no reference junction compensation
(reference junction assumed to be at 0°C).
The
command calibrates temperature channels based on Current Source
*CAL?
SCP and Sense Amplifier SCP setup at the time of execution. If SCP settings are
changed, those channels are no longer calibrated.
This command will cause Error +3000, "illegal while initiated" if trigger is initiated.
Related Commands:
*CAL?
, and
SENSe:REFerence
*RST Condition:
[SENSe:]FUNCtion:VOLTage AUTO,(@100:163)
Link first 20 channels to the K type thermocouple temperature conversion.
SENS:FUNC:TEMP TCOUPLE,K,(@100:119)
Link channel 20 to measure reference temperature using 5K thermistor.
SENS:REF THERM,5000,(@120)
Define Scan List 1 to measure reference channel first then TC channels.
ROUT:SEQ:DEF LIST1,(@120,100:119)
Link channels 21 - 39 to the 10K ohm thermistor temperature conversion,
range is defaulted to autorange.
SENS:FUNC:TEMP THER,10kohm,(@121:139)
<
sub_type
>
values of 2250, 5000, and 10000
sub_type
may be specified in Kohms (kohm).
<
>
applies to E type thermocouples at 800°C and above.
*CAL?
,
OUTPut:CURRent
(for RTDs and Thermistors),
SENSe:REFerence:TEMPerature
must be executed again.
(for Thermocouples)