Loading Tables for
Non-Linear Conversions
Usage Example
Linking Custom
EU Tables
80
Using the HP E1313/E1413
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
DIAGnostic:CUSTom:PIECewise <
command downloads a custom piecewise EU conversion table to the
HP E1413 for each channel specified.
table_block
is a block of 1,024 bytes that define 512 16-bit values.
<
>
SCPI requires that
data header. C-SCPI adds the header for you.
<
table_range
>
specifies the range of input voltage that the table
covers (from –
<
table_range
specify must be within 5% of:
.015625 | .03125 | .0625 | .125 | .25 | .5 | 1 | 2 | 4 | 8 | 16 | 32 | 64.
<
ch_list
>
specifies which channels will have this custom EU table loaded.
Your program puts table constants into array table_block.
DIAG:CUST:PIEC table_block,1,(@124:131) Send table for channels 24-31 to
SENS:FUNC:CUST 1,1,(@124:131)
INITiate then TRIGger module
The
[SENSe:]FUNCtion:CUSTom [<
channels with the custom EU conversion table loaded with the
DIAGnostic:CUSTom:LINear
Two other custom linking commands are available for custom thermocouple
measurements:
[SENSe:]FUNCtion:CUSTom:TCouple <
[SENSe:]FUNCtion:CUSTom:REFerence [<
A full explanation of these custom temperature EU linking commands can
be found starting on page 136.
The
range
parameter selects one of the HP E1413's voltage
<
>
ranges: .0625 | .25 | 1 | 4 | 16. To select a range, simply specify the
range. Specifying
Specifying
AUTO
selects autorange. The default range (no range
parameter specified) is autorange. If an A/D reading is greater than
the
<
table_range
>
an overrange condition will occur.
The
<
type
>
command is for the
command and it specifies the type of thermocouple wire.
If no custom table has been loaded for the channels specified with
SENSe:FUNCtion:CUSTom
command is given.
INITiate
table_range
>,<
table_block
table_block
include the definite length block
<
>
>
to +
<
table_range
HP E1413.
Link custom EU with channels
24-31 and set the 1V A/D range.
range
ch_list
>,](@<
or
DIAGnostic:CUSTom:PIECewise
type
>,[<
and
range
selects the lowest range (.0625VDC).
0
specified with
DIAGnostic:CUSTom:PIECewise
SENSe:FUNCtion:CUSTom:TCouple
, an error will be generated when an
>,(@<
ch_list
>)
>
). The value you
command links
>)
commands.
range
>,](@<
ch_list
>)
>,](@<
ch_list
>)
,