Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
+2129
'Config warning, Sysfail detected'.
+3000
'Illegal while initiated'. Operation must be performed
before
INIT
+3001
'Illegal while continuous'. While
TRIG:SOUR IMM
+3004
'Illegal command. CAL:CONF:RES or
CAL:CONF:VOLT not sent'. Incorrect sequence of
calibration commands. Send
CAL:CONF:VOLT
CAL:VAL:VOLT
+3005
'Illegal command. Send CAL:VAL:RES'. The only
command accepted after a CAL:CONF:RES is a
CAL:VAL:RES command.
+3006
'Illegal command. Send CAL:VAL:VOLT'. The only
command accepted after a CAL:CONF:VOLT is a
CAL:VAL:VOLT command.
+3007
'Invalid signal conditioning module'. The command sent
to an SCP was illegal for its type.
+3008
'Too few channels in scan list'. A Scan List must contain
at least two channels.
+3012
'Trigger too fast'. Scan list not completed before another
trigger event occurs.
+3015
'Channel modifier not permitted here'.
+3019
'TRIG:TIM interval too small for SAMP:TIM interval
and scan list size'.
completion of entire scan list at currently set
interval. See
Reference
+3020
'Input overvoltage'. Calibration relays opened (if
JM2202 not cut) to protect module inputs, and
Questionable Data Status bit 11 set. Execute *RST to
close relays and/or reset status bit.
+3021
'FIFO overflow'. Lets you know that the FIFO buffer has
filled and that one or more readings have been lost.
Usually caused by continuous measurements taken faster
than FIFO was read.
+3026
'Calibration failed'.
or
.
INIT:CONT ON
INIT:CONT ON
.
CAL:CONF:RES
command before
CAL:VAL:RES
command.
interval must allow for
TRIG:TIM
in Chapter 5, the Command
TRIG:TIM
Error Messages
and
or
or
SAMP:TIM
325

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1413c

Table of Contents