Timer Paced Scans - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

Timer Paced Scans

112
Understanding the HP E1313/E1413
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
In this mode the Trigger Timer triggers each execution of a scan list. The
trigger timer (t
) can be set from .1 ms up to 6.5 seconds. Of course, t
p
allow time for all channels to be scanned before the next timer trigger;
> (channel count + 3) ∗ t
t
p
range from 10 µs to 32 ms.
The
TRIGger:COUNt
command controls the number of times any trigger will
be accepted, from
1
to
32767
the module stops scanning until another
executed. Trigger count defaults to 1.
t
p
t
c
t
is the sample time between channels.
c
t
is the pacing time between channel lists.
p
The large arrow is the pacing Trigger Timer. The smaller arrows are sample
triggers generated by the HP E1313/E1413 when stepping through the
channel list. The first large arrow is the arming event.
Example Command Sequence
>,(@101,103,105,107) <n> can be 1 - 4.
ROUT:SEQ:DEF LIST<
n
n
ROUT:SCAN LIST<
>
n
t
SAMP:TIM LIST<
>,<
>
c
TRIG:SOUR TIM
TRIG:TIM:PER <
t
>
p
TRIG:COUN <
count
>
ARM:SOUR <
source
>
INIT:IMM
Starting
ARM event, either hardware or software
Acquisition starts
Stopping Before Trigger Count is Reached
TRIG:SOUR HOLD
TRIG:IMM
or
ABORt
+ 30µs. The sample timer controls t
c
or
INFinite
. When the trigger count is exhausted,
INITiate
-
ARM
<n> can be 1 - 4.
<t
> can be 10µs to 32.768ms.
c
<t
> can be .1ms - 6.5536 s.
p
<count> can be 1 - 32768 or INF.
<source> can be BUS, EXT, HOLD,
IMM, or TTLTrg0 - TTLTrg7.
Stops at end of scan list, still
initiated.
One more scan, now idle.
Stops immediately (must use
ABOR if TRIG:COUNt is INF).
must
p
in the
c
-
TRIGger
sequence is
=Trigger Count
Channel Number
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1413c

Table of Contents