16 -0E 16; Scan Status And Control Register Base - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

Command and Parameter Registers
Note
Scan and Card Control Registers
Scan Status and Control Register
342
Register-Based Programming
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Write only: Commands and their parameters are written to the Command
and Parameter Registers. See " Register-Based Command Reference"
starting on page 356.
Command Register
Address
Base + 08
16
Parameter Register #1
Address
Base + 0A
16
Parameter Register #2
Address
Base + 0C
16
Parameter Register #3
Address
Base + 0E
16
The parameters associated with a command must always be written before
the actual command op code. No command requires more than three
parameter words. Upon a write to the Command Register, the Cmd/Parm
Ready bit is de-asserted (set LOW). It is subsequently asserted after the
control processor has removed the command word and associated
parameters from the Parameter Register buffers and is ready to receive
another command.
Read/Write: This register controls and provides status of the execution of
scan lists. Note that bits 15 through 8 are read-only and are used to report scan
status, and bits 7 through 0 are read/write and are used to control the scan.
Base + 08
15 - 0
Command Op code
15 - 0
Parameter #1 Op code
15 - 0
Parameter #2 Op code
15 - 0
Parameter #3 Op code
-0E
16
16
Base + 10
16
Appendix D