CALibration
CALibration:ZERO?
Comments
Usage
186
HP E1313/E1413 Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
corrects Analog to Digital converter offset for any drift since
CALibration:ZERO?
the last
*CAL?
or
CALibration:ZERO?
•
The
CALibration:ZERO?
the
*CAL?
common command to perform on-line calibration of channels as well
as A/D offset.
performs gain and offset correction of the A/D and each
*CAL?
channel out to the field wiring module connector.
•
Returned Value:
Value
0
-1
The C-SCPI type for this returned value is int16.
•
Executing this command does not alter the module's programmed state (function,
range, etc.).
•
Related Commands:
*CAL?
•
*RST Condition: A/D offset performed.
CAL:ZERO?
command was executed.
command only corrects for A/D offset drift (zero). Use
Meaning
Cal OK
Cal Error
See Error Messages in Appendix B.
Returns 0 or -1.
Further Action
None
Query the Error Queue
(SYSTem:ERRor?)
Chapter 5