Trigger[:Immediate]; Trigger:source; Trigger:source <Source - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

TRIGger[:IMMediate]

Comments
Usage

TRIGger:SOURce

Parameters
HP E1313 Note
Comments
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
causes one trigger when the module is set to the

TRIGger[:IMMediate]

TRIGger:SOURce BUS
or
This command is equivalent to the
GET
bus command.
Related Commands:
TRIGger:SOURce
TRIG:IMM
trig_source
TRIGger:SOURce <
trigger event.
Parameter
Parameter
Name
Type
<
trig_source
>
discrete (string)
Most B-Size mainframes do not support VXIbus TTLTRG lines.
The following table explains the possible choices.
Parameter Value
BUS
EXTernal
HOLD
IMMediate
SCP
TIMer
n
TTLTrg<
>
See the note about the
ARM
The term " Continuous Mode" means that the module is scanning with
TRIGger:SOURce IMMediate
event, the module executes a single Scan List continuously until an
INITiate:CONTinuous OFF
scan rate.
TRIGger:SOURce HOLD
*TRG
common command or the IEEE-488.2
Use TRIGGER to start a measurement scan.
configures the trigger system to respond to the
>
Range of
Values
BUS | EXT | HOLD | IMM | SCP | TIM |
TTLTrg<
Source of Trigger
TRIGger[:IMMediate], *TRG, GET (for HP-IB)
"Trig" signal on terminal module
TRIGger[:IMMediate]
The trigger event is always satisfied.
SCP Trigger Bus (future HP or SCP Breadboard)
The internal trigger timer. See also TRIGger:TIMer:MODE
The VXIbus TTLTRG lines (n=0 through 7)
subsystem on page 266.
and
INITiate:CONTinuous ON
is received. Continuous Mode provides the fastest
HP E1313/E1413 Command Reference
TRIGger
mode.
Default
Units