Route:sequence:define; Route:sequence:define? <Scan_List>,<Mode - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

ROUTe

ROUTe:SEQuence:DEFine?

Parameters
Comments
Usage
224
HP E1313/E1413 Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
ROUTe:SEQuence:DEFine? <
LIST4
, returns either the sequence of channels or the sequence of Channel Data
Modifiers for Scan List
LIST<
of scan list numbers.
Parameter
Parameter
Name
Type
<
scan_list
>
discrete (string)
mode
discrete (string)
<
>
The default for
mode
<
>
When
scan_list
is
<
>
LIST1 - LIST4
ROUTe:SEQuence:DEFine?
When
<
scan_list
>
is
LIST1 - LIST4
ROUTe:SEQuence:DEFine?
current Scan List channels. The meaning of the Channel Data Modifier values are
shown in the following table.
Channel Data
Modifier
1
Perform EU conversion and store result in both FIFO buffer and
Current Value Table.
2*
Leave measurement as voltage and store result in both FIFO and CVT.
3
Perform EU conversion and store result in CVT only.
4*
Leave measurement as voltage and store in CVT only.
5
Perform EU conversion and store result in FIFO only.
6*
Leave measurement as voltage and store in FIFO only.
7*
Leave measurement as voltage and do not store result in either
FIFO or CVT. Use as dummy channel set.
* Limit Checking (CALCulate:LIMit:...) cannot be performed on channels that are not
converted to Engineering Units.
Returned Value: Returned values are as a comma separated list.
Related Commands:
ROUTe:SEQuence:DEFine
*RST Condition: Scan List 1 =
Scan Lists 2 through 4 have no channels assigned to them,
assigned to it.
ROUT:SEQ:DEF? LIST2, CHAN
Enter comma separated list of channels
ROUT:SEQ:DEF? LISTL
Enter comma separated list of Scan Lists
scan_list
mode
>[,<
>]
n
>
. When
<
scan_list
>
Range of
Values
LIST1 | LIST2 | LIST3 | LIST4 | LISTL
CHANnel | MODifier
is
.
CHAN
and
mode
is
<
>
CHAN
returns the sequence of channels assigned to
and
<
mode
>
is
returns the Channel Data Modifiers assigned to the
Description
,
(@100:163)
(Channel Data Modifiers are all 1),
Query for channel sequence in LIST2.
Query for scan list sequence in LISTL.
When
scan_list
is
<
>
LIST1
is
LISTL
, returns the sequence
Default
Units
none
none
,
<
scan_list
MOD,
ROUTe:SCAN
LISTL
as no Scan Lists