Cls; Dmc <Name>,<Cmd_Data; Emc; Emc <Enable - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

Common Command Reference
Note

*CLS

*DMC <name>,<cmd_data>

*EMC

*EMC?

276
HP E1313/E1413 Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
If Open Transducer Detect (OTD) is enabled when
will disable OTD, wait 1 minute to allow channels to settle, perform the calibration,
and then re-enable OTD. If your program turns off OTD before executing
should also wait 1 minute for settling.
Clear Status Command. The
(Standard Event Status Event Register, Standard Operation Status Event Register,
Questionable Data Event Register) and the instrument's error queue. This clears the
corresponding summary bits (bits 3, 5, and 7) in the Status Byte Register.
does not affect the enable bits in any of the status register groups. (The SCPI
command
STATus:PRESet
Data Enable Registers.)

*CLS

command) and the Operation Complete Query function (
Define Macro Command. Assigns one, or a sequence of commands to a named
macro.
The command sequence may be composed of SCPI and/or Common Commands.
<
name
>
may be the same as a SCPI command, but may not be the same as a
common command. When a SCPI named macro is executed, the macro rather than
the SCPI command is executed. To regain the function of the SCPI command,
execute *EMC 0.
<
cmd_data
>
is sent as arbitrary block program data. For more information see
page 156.
Enable Macro Command. When
When
<
enable
>
is zero, macros are disabled.
Enable Macro Query. Returns either
or
0
(macros are disabled).
*CAL?
command clears all status event registers
*CLS
does clear the Operation Status Enable and Questionable
disables the Operation Complete function (
enable
is non-zero, macros are enabled.
<
>
1
(macros are enabled),
is executed, the module
, it
*CAL?
*CLS
*OPC
*OPC?
command).
Chapter 5

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1413c

Table of Contents