[SENSe:]FUNCtion:STRain:FBENding

[SENSe:]FUNCtion:STRain:FBPoisson
[SENSe:]FUNCtion:STRain:FPOisson
[SENSe:]FUNCtion:STRain:HBENding
[SENSe:]FUNCtion:STRain:HPOisson
[SENSe:]FUNCtion:STRain[:QUARter]
Parameters
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
*RST Condition:
SENSe:FUNCtion:VOLTage AUTO,(@100:163)
FUNC:RES 30ua,(@100,105,107)
Note on Syntax: Although the strain function is comprised of six separate SCPI
commands, the only difference between them is the bridge type they specify to the
strain EU conversion algorithm.
[SENSe:]FUNCtion:STRain:<
EU conversion with the channels specified by
See "Linking Channels to EU Conversion" in Chapter 3 for more information.
<
bridge_type
>
is not a parameter but is part of the command syntax. The following
table relates the command syntax to bridge type. See the User's Manual for the
optional Strain SCP for bridge schematics and field wiring information.
Command
:FBENding
:FBPoisson
:FPOisson
:HBENding
:HPOisson
[:QUARter]
Parameter
Parameter
Name
Type
range
<
>
numeric (float32)
ch_list
<
>
channel list (string)
Set channels 0, 5, and 7 to convert voltage
to resistance assuming current source set to
30 µA use autorange (default).
bridge_type
> [<
range
>,](@<
<
ch_list
>
to measure the bridge voltage.
Bridge Type
Full Bending Bridge
Full Bending Poisson Bridge
Full Poisson Bridge
Half Bending Bridge
Half Poisson Bridge
Quarter Bridge (default)
Range of
Values
.0625 | .25 | 1 | 4 | 16
(See Comments)
100 - 163
HP E1313/E1413 Command Reference
[SENSe:]
ch_list
>)
links the strain
Default
Units
VDC
none