SYSTem
Subsystem Syntax
SYSTem:CTYPe?
Parameters
Comments
Usage
SYSTem:ERRor?
Comments
264
HP E1313/E1413 Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The
SYSTem
subsystem is used to query for error messages, types of Signal
Conditioning Plug-ons (SCPs), and the SCPI version currently implemented.
SYSTem
:CTYPe? (@<
:ERRor?
:VERSion?
channel
SYSTem:CTYPe? (@<
Conditioning Plug-on installed at the specified channel.
Parameter
Parameter
Name
Type
<
channel
>
channel list (string)
• <
channel
>
must specify a single channel only.
•
Returned Value: An example of the response string format is:
HEWLETT-PACKARD,E1413 Model
SCP,0,0
•
The C-SCPI type returned is string. For specific response string, refer to the
appropriate SCP manual. If
installed, the module returns the response string:
0,No SCP at this Address,0,0
SYST:CTYP? (@100)
returns the latest error entered into the Error Queue.
SYSTem:ERRor?
• SYSTem:ERRor?
returns one error message from the Error Queue (returned error
is removed from queue). To return all errors in the queue, repeatedly execute
SYSTem:ERRor?
until the error message string = +0, " No error" .
•
Returned Value: Errors are returned in the form:
±<error number>, " <error message string>"
•
RST Condition: Error Queue is empty.
channel
>)
returns the identification of the Signal
>)
Range of
Values
100 - 163
model number and description
<
<
channel
>
specifies a position where no SCP is
Return SCP type install at channel 0.
SYSTem
Default
Units
none
>
Chapter 5