HP E1313A Manual page 389

High speed a/d module
Table of Contents

Advertisement

388
Using HP VEE with the HP E1313/E1413
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Direct access to the HP E1313/E1413 CVT table is configured by
selecting:
I/O
Instrument
Highlight the HP E1313/E1413 in your system
Edit instrument
A24/A32 Space Config...
Then set the following fields:
Byte Access D16
Word Access D16
Long Word Access (D32) D32 Access
Name Offset Format
CVT
0
Real 32
The HP E1313/E1413 current value table is located in A24 address
space. Each channel is a 32-bit real number with Channel 0 at offset
0, channel 1 is at offset 4, and so on.
If your hardware supports it, access to both the FIFO and the CVT
should use D32 Access as it is twice as fast as D16 access. If using a
VXLINK (HP E1383A or HP E1483A) as the interface or if using an
HP E1413A, then only D16 access is supported. To use only D16
Access set:
Long Word Access (D32)
NONE
D16 Access
always selected
Mode
READ
NONE
Appendix E