Example Command Sequence
Card 1
n
>,(@101,103,105,107) <n> can be 1 - 4.
ROUT:SEQ:DEF LIST<
ROUT:SCAN LIST<
n
>
SAMP:TIM LIST<
n>,<t
>
c
TRIG:SOUR TIM
t
TRIG:TIM:PER <
>
p
TRIG:TIM:MODE SYNC
TRIG:COUN 2
ARM:SOUR <
source
>
OUTP:TTLT<
n
>:SOUR FTR
OUTP:TTLT<
n
>:STAT ON
INIT:CONT ON
Card 2
n
>,(@101,103,105,107) <n> can be 1 - 4.
ROUT:SEQ:DEF LIST<
n
ROUT:SCAN LIST<
>
n
t
SAMP:TIM LIST<
>,<
>
c
TRIG:SOUR TTLTrg<
n
>
TRIG:COUN <
count
>
INIT:IMM
Starting
ARM event, either hardware or software
Acquisition starts
Stopping Before Trigger Count is Reached
INIT:CONT OFF
TRIG:SOUR HOLD
TRIG:IMM
or
ABOR
Understanding the HP E1313/E1413
<t
> can be 10µs to 32.768ms.
c
<t
> can be .1ms - 6.5536 s.
p
Keep timer running during
INIT:CONT ON.
<source> can be BUS, EXT,
HOLD, IMM, or TTLTrg0 -
TTLTrg7.
1st trig will source TTLTrg.
Enable to drive selected TTLTrg
line. <n> can be 1 through 7.
<t
> can be 10µs to 32.768ms.
c
Selects TTLTrg trigger source.
<count> can be 1 - 32768 or INF.
To master card.
Stops at end of scan list, still
initiated.
One more scan, now idle.
Stops immediately (must use
ABOR if TRIG:COUNt is INF).
119