Triggering And Scanning Modes; Continuous (Free-Run) Mode - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

Triggering and Scanning Modes

Continuous
(Free-Run) Mode
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The following section shows the various ways the HP E1313/E1413 can
control channel scanning, and how to select the trigger mode to accomplish it.
The Continuous Mode provides the fastest scanning. The time between
channel measurements is controlled by the trigger timer interval and is
consistent from channel-to-channel as well as from the last channel in the
scan list back to the first channel in the list when the scan list is re-executed.
ARM
t
is the sample time between channels.
c
The large arrow is the arming event or signal. The smaller arrows
are sample triggers generated by the HP E1313/E1413 when
stepping through the channel list.
Example Command Sequence
n
>,(@101,103,105,107) <n> can be 1 - 4.
ROUT:SEQ:DEF LIST<
ROUT:SCAN LIST<
n
>
SAMP:TIM LIST<n>,<t
>
c
TRIG:SOUR IMM
source
ARM:SOUR <
>
INIT:CONT ON
Starting
ARM event, either hardware or software
Acquisition starts
Stopping
INIT:CONT OFF
Understanding the HP E1313/E1413
t
c
<t
> can be 10µs to 32.768ms.
c
<source> can be BUS, EXT,
HOLD, IMM, or TTLTrg0 -
TTLTrg7.
Stops at end of scan list.
Channel Number
111

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1413c

Table of Contents