[SENSe:]DATA:FIFO:MODE

Parameters
Comments
Usage

[SENSe:]DATA:FIFO:MODE?

Comments
Usage
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe:]DATA:FIFO:MODE <
reading memory.
Parameter
Parameter
Name
Type
<
mode
>
discrete (string)
In
BLOCk
(ing) mode, if the FIFO becomes full and measurements are still being
made, the new readings are discarded. In
full and measurements are still being made, new readings overwrite the oldest
readings. In this situation, the FIFO contains only the latest 65,024 readings. In
both modes Error +3021, " FIFO Overflow" is generated to let you know that
measurements have been lost.
This command will cause Error +3000, "illegal while initiated" if trigger is initiated.
Related Commands:
[SENSe:]DATA:FIFO:MODE?
[SENSe:]DATA:FIFO[:ALL]?
[SENSe:]DATA:FIFO:PART?
*RST Condition:
SENSe:DATA:FIFO:MODE BLOCk
SENS:DATA:FIFO:MODE OVERWRITE
SENS:DATA:FIFO:MODE BLOCK
[SENSe:]DATA:FIFO:MODE?
Returned Value: String value either
type returned is string.
Related Commands:
SENSe:DATA:FIFO:MODE
SENSe:DATA:FIFO:MODE?
mode
sets the mode of operation for the FIFO
>
Range of
Values
BLOCk | OVERwrite
OVERwrite
,
,
[SENSe:]DATA:FIFO:HALF?
,
[SENSe:]DATA:FIFO:COUNt?
Set FIFO to overwrite mode.
Set FIFO to block mode.
returns the currently set FIFO mode.
BLOCK
or
OVERWRITE
Enter statement returns either BLOCK or
OVERWRITE.
HP E1313/E1413 Command Reference
[SENSe:]
Default
Units
none
Mode, if the FIFO becomes
,
. The C-SCPI