parameter must be within ± 5% of one of the 5 following values:
The
range
<
>
.0625 VDC, .25 VDC, 1 VDC, 4 VDC, 16 VDC.
millivolts (mv).
• CALibration:CONFigure:RESistance
must be sent before
CALibration:VALue:RESistance
commands.
The only
command accepted after
CALibration
CALibration:VALue:VOLTage
The
CALibration:STORe
in non-volatile memory.
Perform
ZERO
before
FSCale
The
output voltage of the calibration source will be greater than 90% of
FSCale
the nominal value for each range, except the 16 V range where the output is 10 V.
This command will cause Error +3000, "illegal while initiated" if trigger is initiated.
Related Commands:
CALibration:STORe ADC
CAL:CONF:VOLT .0625,ZERO
*OPC?
Now measure voltage with external DMM.
measured value
CAL:VAL:VOLT <
CAL:CONF:VOLT .0625,FSCale
*OPC?
Now measure voltage with external DMM.
CAL:VAL:VOLT <
measured value
Repeat these activities for each of 4 other ranges (.25, 1, 4, 16).
.
.
.
CAL:STOR ADC
<
or
CALibration:CONFigure:VOLTage
or
CALibration:CONFigure:VOLTage
.
command may be used to store the calibration constants
or Error +3050 will be generated.
,
CALibration:VALue:VOLTage
Connect zero voltage reference to
Calibration Bus.
Must wait for CAL:CONF:VOLT to complete.
>
Inform HP E1413 of measured value.
Connect full-scale voltage reference to
Calibration Bus.
Must wait for CAL:CONF:VOLT to complete.
>
Inform HP E1413 of measured value.
Store calibration constants in non-volatile
memory (use only at end of complete cal
sequence).
range
may be specified in
>
commands
CALibration:VALue:VOLTage
Chapter 5