[SENSe:]DATA:FIFO:RESet

Comments
Usage

[SENSe:]FILTer[:LPASs][:STATe]

Parameters
Comments
Note
Usage
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

[SENSe:]DATA:FIFO:RESet

reset to 0.
This command will cause Error +3000, "illegal while initiated" if trigger is initiated.
Related Commands:
[SENSe:]DATA:FIFO:...
*RST Condition:
SENSe:DATA:FIFO:RESet
SENS:DATA:FIFO:RES
[SENSe:]FILTer[:LPASs][:STATe] <
low-pass filter.
Parameter
Parameter
Name
Type
enable
<
>
discrete (string)
The A/D low-pass filter when enabled affects all channels.
The filter's cut-off frequency is 12 KHz.
Autorange (the default voltage range) is not allowed while the A/D low-pass filter
is enabled. If channels contained in the current Scan List specify autorange, while
the A/D filter is
ON
an error will be generated at
This command will cause Error +3000, "illegal while initiated" if trigger is initiated.
When the A/D filter is
ON
is 145 µsec.
for the current scan list is less than 145 µsec and the A/D filter is
If
SAMPle:TIMer
, an error will be generated at
ON
Related Commands:
[SENSe:]FILTer[:LPASs][:STATe]?
*RST Condition:
SENSe:FILTer:LPASs:STATe OFF
SENS:FILT:LPAS:STAT ON
clears the FIFO of readings. The FIFO counter is
commands.
Clear the FIFO.
enable
>)
enables or disables the A/D's
Range of
Values
ON | OFF
INITiate:IMMediate
, the minimum allowable value for
time.
INIT
Enable A/D low-pass filter.
HP E1313/E1413 Command Reference
[SENSe:]
Default
Units
none
time.
SAMPle:TIMer
235