Figure 3-9. Center Thermistor Measurements on Multiple Panels
96
Using the HP E1313/E1413
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
If two Terminal Panels are used, each panel must be connected as above so
that both panels provide reference temperature measurements. These
connections are shown in Figure 3-9.
Two HP E1586 Terminal Panels
Measuring One Thermistor on Each Panel
on Reference Channels 100, 132
The following example uses two Terminal Panels to measure type K
thermocouples.
SENS:REF THER,5000,1,(@100,132)
Measures reference temperature measurements on channels 100 and 132.
SENS:FUNC:TEMP TC,K,.06,(@101:131,133:163)
Defines channels for temperature measurements.
ROUT:SEQ:DEF (@100,101:131,132,133:163)
Defines the scan list.