Notes
328
Error Messages
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
These are the only tests where the user should
troubleshoot a problem. Other tests which fail must be
handled by an HP repair facility. Make sure you record
the values found in the FIFO for HP Service.
Refer to the Command Reference under
of module functions tested.
1. HP E1413/E1313 C-SCPI driver for MS-DOS
versions of *TST. The default version is an abbreviated self test that
executes only the Digital Tests. By loading an additional object file,
you can execute the full self-test. See the documentation that comes
with the HP E1413 C-SCPI driver for MS-DOS
2. During the first 5 minutes after power is applied,
Allow the module to warm-up before executing
+3053
'Corrupt on board Flash memory'.
+3056
'Custom EU not loaded'. May have erased custom
EU conversion table with
channel with standard EU after loading custom EU, this
erases the custom EU for this channel. Reload custom EU
table using
+3058
'Hardware does not have D32, S/H, or new trigger
capabilities'. Module's serial number is earlier than
3313A00530.
+3059
'DSP firmware does not have LISTL, LIMIT, or
AVERAGE capability'. Firmware (stored in Flash
Memory) must be updated to version A.03.00 or later.
+3060
'Lower limit exceeds upper limit on one or more
channels'.
channel's
+3061
'Cannot change scan list while running with AVERAGE
on'
+3062
'LISTL not allowed with AVERAGE on'
+3063
'Duplicate channels in list not allowed with AVERAGE
on'
+3064
'TRIG:COUNT must be >= CALC:AVER:COUNT if
INIT:CONT not on'
®
®
*RST
. May have linked
or
DIAG:CUST:LIN
DIAG:CUST:PIEC
CALC:LIM:LOW:DATA
set higher than
CAL:LIM:UPP:DATA
*TST?
for a list
implements two
.
may fail.
*TST?
*TST?.
.