Measuring Using the
To measure all three thermistors, select three HP E1313/E1413 channels as
Left, Center, and Right
reference channels. The recommended method is to use thermistor 1, the
left thermistor (HI-T1S and LO-T1S), as a reference for channels 0 - 7
Thermistors
(or 32 - 39), thermistor 2, the center thermistor (HI-T2S and LO-T2S), for
channels 8 - 23 (or 40 - 47), and thermistor 3, the right thermistor (HI-T3S
and LO-T3S), for channels 24 - 31 (or 56 - 63). Connect the reference
channel's HI and LO terminals to the appropriate thermistor terminals
(e.g., channel 100 HI and LO terminals to thermistor 1 HI-T1S and LO-T1S
terminals, respectively). These connections are shown in Figure 3-10.
One HP E1586 Terminal Panel
Measuring Three Thermistors
on Reference Channels 100, 101, 102
Figure 3-10. Left, Center, and Right Thermistor Measurements
on a Single Panel
If two Terminal Panels are used, each panel must be connected as above so
that both panels provide reference temperature measurements. These
connections are shown in Figure 3-11.