HP E1313A Manual page 127

High speed a/d module
Table of Contents

Advertisement

126
Understanding the HP E1313/E1413
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
If you wanted the " FIFO half full" and " scan complete" conditions to be
reported you would execute:
STAT:OPER:ENAB 1280
Standard Event Group Examples
If you only wanted the " query error" , " execution error" , and " command
error" conditions to be reported by the ESB bit (bit 5) of the Status byte,
you would execute:
*ESE 52
Operation Group Example
Suppose that you wanted the module to report via the Status system when it
had completed executing
*CAL?
Operation Condition Register goes to 1 when
returns to 0 when
*CAL?
is complete. In order to record only the negative
transition of this bit in the Status Operation Event Register you would send:
STAT:OPER:PTR 32766
STAT:OPER:NTR 1
Now when
completes and Operation Condition Register bit zero goes
*CAL?
from 1 to 0, Operation Event Register bit zero will become a 1.
1280 = decimal sum of values for
bits 10 and 8.
52 = decimal sum of values for
bits 2, 4, and 5.
. The " Calibrating" bit (bit 0) in the
is executing and
*CAL?
All ones in Positive Transition
Filter Register except bit 0 = 0.
All zeros in Negative Transition
Filter Register except bit 0 = 1.
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1413c

Table of Contents